3.

Installation

3.1 Site requirements

Make sure to read and follow all
safety instructions in chapter 2.3.2
'Site requirements'!

3.1.1 General site requirements

• No direct sunlight.
• Electrical power supply within distance
(length of power cord = approx. 4 meters -
do not use extension cords!).
• No draft (caused by air conditioning etc.).
• Even floor surface.
• Practically vibration-free floor.
• Handwheel easily accessible.
• Room temperature constantly below +22 °C.
• Relative humidity of air maximum 60 %.
High ambient temperature and/or high
air
humidity
instrument cooling performance!

3.1.2 Electrical connections

Make sure to read and follow all
safety instructions in chapter 2.3.3
'Electrical connections'!
16
negatively
affect

3.2 Unpacking and installation

Unpacking instructions are always located in a
transparent protective envelope on the outside
of the instrument shipping crate.
Make sure to read and follow all
safety
instructions
chapter 2.3.1 'Transport' and on the
unpacking instructions!

3.2.1 Repacking

We recommend to keep the original shipping
crate and the unpacking instructions for the
Leica CM3050 S.
For repacking, proceed as per unpacking in-
structions, in reverse order.
Available models
• Basic instrument
- with sectioning motor
- without specimen cooling
• Basic instrument
- without sectioning motor
- with specimen cooling
• Basic instrument
- with sectioning motor
- with specimen cooling
Instruction Manual V1.2 - 11/2006