Setup Example Using The Sfp - Keysight M8195A User Manual

Arbitrary waveform generator revision 2
Hide thumbs Also See for M8195A:
Table of Contents

Advertisement

4.3.2

Setup example using the SFP

Keysight M8195A Revision 2 – Arbitrary Waveform Generator User's Guide
Select the waveform segment, that will be executed immediately after starting
the signal generation.
:TRAC:SEL 1
Start signal generation.
:INIT:IMM
Waveform update and switch operation in a loop until stopped by :ABOR command:
Reload data into next segment. The <segment_id> is either 1 or 2.
:TRAC:DATA <segment_id>,0,#41280<data-bytes>
Dynamically switch to reloaded segment The <sequence_table_index> is either
0 or 1.
:STAB:DYN:SEL <sequence_table_index>
The following section shows the setup using the SFP.
Preparation:
Set the continuous mode in the Trigger Tab.
Set sequencing mode to ARBitrary in the Sequence/Control Tab.
Set dynamic mode in the Sequence/Control Tab.
Create two waveform segments in the Standard Waveform Tab.
Create two sequence table entries referring to the waveform segments in the
Sequence/Control Tab. Use the "Select Init Segment" field to select the
waveform segment, that will be executed immediately after starting the signal
generation. Send the sequence table to the M8195A.
Start signal generation by pressing the Run/Stop button.
Streaming 4
147

Hide quick links:

Advertisement

Table of Contents
loading