3.6.1.1.2
Asynchronous Triggering
Keysight M8195A Revision 2 – Arbitrary Waveform Generator User's Guide
In synchronous trigger mode the incoming trigger and event signals are sampled with
the SYNC clock which is the DAC sample rate divided by 256 and the input signals need
to be provided synchronous to the SYNC clock to get a precise output signal.
When using the asynchronous mode, the trigger and event input signals are sampled
with a clock that is the DAC sampling rate divided by 8. This provides a more precise
trigger/event to output latency without the need of providing the inputs synchronous to
any reference.