Setting Upper and Lower Respiratory Rate (RR/BR) Limits
Initially, the RR/BR Alarm menu specifies that RR/BR alarms are enabled (
(
), and displays the default Upper and Lower respiratory limits. The upper and lower limits
OFF
can be
ON
adult, pediatric, and neonate patients in Breaths/Minute, and gives the range in which you can
set these limits:
Patient Type
Adult
Pediatric
Neonate

Using the CO2 Parameter Control Panel

Use the navigation keys to highlight and select the CO2 numeric display to display the
CO2 Parameter Control Panel:
Figure 10-2 CO2 Parameter Control Menu
On the CO2 Parameter Control Panel, you can select a CO
FiCO2). On the selected alarm settings menu, you can enable/disable alarms and set alarm
limits (as described previously). You can also set the CO
10-10
CO2
or
(default is
). The following table lists the default Respiratory limits for
OFF
OFF
Respiration Rate Default
Lower: 3 BPM
Upper: 50 BPM
Lower: 3 BPM
Upper: 50 BPM
Lower: 12 BPM
Upper: 80 BPM
CO2
CO2 Scale
EtCO2 Alarm
0 to 30 mmHg
RR/BR Alarm
0 to 60 mmHg
0 to 100 mmHg
FiCO2 Alarm
0 to 150 mmHg
No Breath Alarm
CO2/Resp Sweep Speed
www.zoll.com
Respiration Rate Range
Lower: 0 to 145 BPM
Upper: 5 to 150 BPM
Lower: 0 to 145 BPM
Upper: 5 to 150 BPM
Lower: 0 to 145 BPM
Upper: 5 to 150 BPM
Lower
Upper
28
50
10
30
0
10
30 sec
12.5 mm/s
alarm (EtCO2, Breath Rate, or
2
sweep speed.
2
) or disabled
ON
9650-0802-06 Rev. A