ZOLL Propaq MD Operator's Manual page 77

Portable defibrillator
Hide thumbs Also See for Propaq MD:
Table of Contents

Advertisement

SpO2
(%)
EtCO2
(mmHg)
FiCO2
(mmHg)
Temp
(°C)
Temp
(°F)
9650-0802-06 Rev. A
Entire range
Limit = 100 (Adult and
Pediatric)
Limit = Numeric + 5 (Neo-
nate)
Entire range
Limit = Numeric + 10
Entire range
Limit = Numeric + 5
Entire range
Limit = Numeric + 0.5
Entire range
Limit = Numeric + 0.9
Propaq MD Operator's Guide
Alarm Options
Limit = Numeric – 5
Limit = Numeric - 5 mmHg
unless the numeric falls
below the lower alarm limit
range, in which case Stat
Set sets the lower limit to
15 mmHg.
N/A
Limit = Numeric – 0.5
Limit = Numeric – 0.9
5-9

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents