Applying Electrodes to the Patient

Depending on local usage, ECG lead wires are marked with certain labels. Refer to the
following table for labels and color codes for the different lead sets
Location
Right Arm
Right Leg
1
American Heart Association
2
International Electrotechnical Commission
Patients should be in a resting, supine position when performing 12-Lead ECG monitoring.
ZOLL Medical Corporation recommends placing the limb electrodes anywhere along the
ankles and wrists.
When it is difficult for the patient to remain motionless due to shivering, muscle tremors, or
transport vehicle movement, place limb electrodes on the patient's thorax for better results.
(Refer to the following two diagrams for limb electrode placement).
9650-0802-06 Rev. A
Left Arm
Left Leg
Chest
Chest
Chest
Chest
Chest
Chest
Propaq MD Operator's Guide
12-Lead ECG Monitoring Setup
1
AHA
Labels
RA (white)
LA (black)
RL (green)
LL (red)
V1
V2
V3
V4
V5
V6
2
IEC
Labels
R (red)
L (yellow)
N (black)
F (green)
C1
C2
C3
C4
C5
C6
13-5