NIBP System Messages

When monitoring NIBP, the Propaq MD unit may display the following messages:
System Message
READING IN PROGRESS
READING STOPPED
READING FAILED
ARTIFACT
NEONATE ALERT
AIR LEAK
KINKED HOSE
NIBP DISABLED
NIBP CALIBRATION CHECK DUE The annual calibration check is due. You can
9650-0802-06 Rev. A
Cause
The unit is taking an NIBP measurement and
functioning normally.
The unit has stopped an NIBP measurement,
because the operator has pressed the NIBP button
and cancelled the measurement.
The patient's pulse is too weak to for NIBP
measurement, or the cuff requires adjustment.
Excessive artifact is preventing the NIBP
measurement.
The unit has detected a neonate cuff when in Adult
mode. Replace cuff or correct patient mode, as
appropriate.
A major air leak is preventing cuff inflation. Check
hose and cuff connections, replace a defective hose
or cuff, as necessary, and reattempt NIBP
measurement.
An air obstruction is preventing the correct operation
of the unit. Check hose for kinks. Eliminate kinks or
replace hose, if necessary.
A system error has occurred, and the Propaq MD
unit should be serviced.
continue to use the device, however, we recommend
that you contact the ZOLL Technical Service
Department or a ZOLL authorized service
representative as soon as possible to perform the
recommended annual NIBP calibration check.
Propaq MD Operator's Guide
NIBP System Messages
8-15