Defibrillation - ZOLL Propaq MD Operator's Manual

Portable defibrillator
Hide thumbs Also See for Propaq MD:
Table of Contents

Advertisement

Defibrillation

The ZOLL Propaq MD can deliver 200 joules of electrical energy. If this electrical energy is not
discharged properly, as described in the this manual, the electrical energy could cause personal
injury or death to the operator or bystander.
To avoid possible damage to the Propaq MD unit, turn off pacing before defibrillating the
patient with a second defibrillator.
After a synchronized cardioversion, the SYNC mode may be cleared after each shock or
disarm. The user may have to reselect (press) the SYNC button after each synchronized
cardioversion shock performed on a patient. In Defib/Pacer Default settings in the Supervisor
Setup menu, the Propaq MD can be configured to remain in the SYNC mode after each
synchronized cardioversion.
Synchronized cardioversion can be performed in the paddle monitoring mode. However, it is
possible that artifact can be produced by the moving paddles, which could cause the
defibrillator to trigger on the artifact. It is recommended that monitoring in leads I, II or III be
used during synchronized cardioversion. Paddle monitoring should not be used for elective
cardioversions procedures.
To avoid stress to the defibrillator or the tester, never attempt to repeatedly charge and
discharge the defibrillator in rapid succession. If a need for repetitive testing arises, allow a
waiting period of at least 2 minutes after every third discharge.
In the SYNC mode, the defibrillator does not discharge without a command signal (R-wave
detection) from the ECG monitor indicated by a SYNC marker on the trace and a flashing
SYNC indicator.
If conductive gel forms a continuous path between the defibrillator electrodes, delivered energy
may be dramatically reduced to zero. In this case, reposition the electrodes to eliminate the
shunting path before attempting additional shocks.
Improper defibrillation technique can cause skin burns. To limit possible skin burns, use only
ZOLL defibrillation gel on paddles, ensure the gel covers the entire paddle surface and press
firmly against patient's chest.
If a new energy level is selected after the CHARGE button is pushed and while the defibrillator
is charging or charged, the defibrillator will disarm. The CHARGE button will need to be
pressed again to charge to the new energy level.
Prior to defibrillation, disconnect from the patient any medical electronic device that is not
labeled "defibrillation protected."
Before charging the defibrillator, verify that the energy selected on the display is the desired
output.
Defibrillation takes priority over external pacing. Should the defibrillator be charged during the
administration of external pacing, the pacer turns off and the defibrillator charges to the
selected energy.
9650-0802-06 Rev. A
Propaq MD Operator's Guide
Warnings
1-15