Test gas for
validation
Validation cell with
60 mm length
Validation cell with
100 mm length
Two validation cells
with 160 mm length
LS4000, LS4060 DIODE LASER ANALYZERS | OI/LS4000/O2-EN REV. A
Proceed as follows:
Step
Procedure
1
Connect a test gas to the test gas input.
2
Connect a discharge line to the test gas output.
The validation is used to check the functional capability of the analyzer.
For the test gas, a gas mixture must be used that contains the measured
component in a concentration suited to generating a significant change in the
display of measured value (typically approx. 80 % of the upper range value, for
approximate values refer to the following tables).
0 to 1 vol. %
MR
OPL
c
c
test gas
total
1 m
14.6
0.8
5 m
73.1
0.8
10 m
100.0
0.5
15 m
100.0
0.4
20 m
100.0
0.3
MR
0 to 1 vol. %
OPL
c
c
test gas
total
1 m
8.5
0.8
5 m
42.2
0.8
10 m
84.4
0.8
15 m
100.0
0.6
20 m
100.0
0.5
MR
0 to 1 vol. %
OPL
c
c
test gas
total
1 m
5.2
0.8
5 m
25.9
0.8
10 m
51.7
0.8
15 m
77.5
0.8
20 m
100.0
0.8
MR
Measuring range
OPL
Optical path length
c
Concentration of the test gas to be emitted for validation in vol. %
test gas
c
Ensuing total concentration in the process in vol. %
total
0 to 10 vol. %
0 to 25 vol. %
c
c
c
test gas
total
test gas
100.0
5.5
100.0
100.0
1.1
100.0
100.0
0.5
100.0
100.0
0.4
100.0
100.0
0.3
100.0
0 to 10 vol. %
0 to 25 vol. %
c
c
c
test gas
total
test gas
84.4
8.0
100.0
100.0
1.9
100.0
100.0
0.9
100.0
100.0
0.6
100.0
100.0
0.5
100.0
0 to 10 vol. %
0 to 25 vol. %
c
c
c
test gas
total
test gas
51.7
8.0
100.0
100.0
3.1
100.0
100.0
1.5
100.0
100.0
1.0
100.0
100.0
0.8
100.0
c
total
5.5
1.1
0.5
0.4
0.3
c
total
9.5
1.9
0.9
0.6
0.5
c
total
15.5
3.1
1.5
1.0
0.8
191