ABB LS4000 Operating Instructions Manual page 163

Diode laser analyzers
Hide thumbs Also See for LS4000:
Table of Contents

Advertisement

Completing the fine
alignment
Ready to perform
measurements
LS4000, LS4060 DIODE LASER ANALYZERS | OI/LS4000/O2-EN REV. A
Proceed as follows:
Step
Procedure
1
Unscrew the 4 stud screws until the ends of the screws no longer
stick out of the holes.
2
While observing the value "Absolute transmission" in the "Align-
ment" menu, align the purging flange by adjusting the 4 fastening
nuts. Adjust the purging flange so that the highest possible value
is displayed for "Absolute transmission".
Proceed as follows:
Step
Procedure
1
Screw in the 4 stud screws as far as they will go so that they se-
cure the setting.
2
Click on "Done" in the "Alignment" menu.
3
If you do not want to perform any further tasks on the instrument
software, disconnect the RJ45 adapter board from the junction
box.
4
To guarantee explosion protection, close the junction box.
The analyzer is set to the process parameters specified by the user when
delivered. If these specifications correspond to the conditions that actually
apply to the process (temperature, pressure, optical path length etc.), then
the analyzer can be put directly into operation after fine alignment.