LS4000, LS4060 DIODE LASER ANALYZERS | OI/LS4000/O2-EN REV. A
Align the purging flanges as follows:
Step
Procedure
1
Unscrew the 4 stud screws 1 until the ends of the screws no longer
protrude from the holes.
2
Switch on the laser pointer.
3
Align the purging flange by adjusting the 4 fastening nuts 2 until
the laser beam hits the center of the focusing screen.
4
Turn the 4 stud screws 1 until they reach a stop, so that they fix
the setting.
5
Switch off the laser pointer.
6
Disassemble the laser pointer and the focusing screen.
The clamps for fastening the laser alignment tool must not be
used for mounting the transmitter unit or receiver unit.
71