When all points have been measured, you can display the maximum, average, and minimum values. To do this,
press
three times to cycle through the maximum, average, and minimum values respectively.
To exit MAP mode, long-press
For each map, a file is created containing all measurement points. This file can then be downloaded and
analyzed.

3.2. Recording Measurements

You can start and stop a recording session on the instrument. Recorded data is stored in the instrument's
memory, and can be downloaded and viewed on a computer running the DataView Data Logger Control Panel.
You can record data by pressing the
A short press (MEM) records the current measurement(s) and date.
A long press (REC) starts the recording session. While the recording is in progress, the symbol REC
appears at the top of the display. A second long press of
the instrument is recording, a short press of
To schedule recording sessions, and download and view recorded data, see §4.

3.3. Errors

The instrument detects errors and displays them in the form Er.XX:
Er.01
Hardware malfunction detected. The instrument must be sent in for repair.
Er.02
Error in internal memory. Format the memory.
Er.03
Hardware malfunction detected. The instrument must be sent in for repair.
Er.10
The instrument has not been correctly adjusted. The instrument must be sent in for calibration.
Er.11
The firmware is incompatible with the instrument. Install the correct firmware (see §6.3).
Er.12
The firmware version is incompatible with the instrument. Reload the previous firmware version.
Er.13
Recording scheduling error. Ensure that the instrument's time and the time of the DataView Data
Logger Control Panel are the same (see §2.3).