FOR BEST ACCURACY
5
All tests that begin with 570 are only for use with the eXact iDip
570
®
photometer.
Test procedures can vary from test to test. Visit
exactidip.com
or
exactidip570.com
to read full instructions and watch instructional video for
each test.
The meter has a default 5 minute automatic shutoff timer. You can change
the duration in the 'Settings' menu located within the eXact iDip
app.
®
Always fi ll the cell to capacity (4mL).
Test immediately after fi lling the cell with the water sample.
Due to the strip slitting process, you may fi nd one or two strips that are
noticeably smaller or larger in width than the normal strips in the bottle.
These should be discarded. Using these strips may give unreliable results.
When testing pH, it is recommended to run the pH test prior to running
Chlorine. If you choose to run the pH after Chlorine, ensure you thoroughly
clean the cell with water and the brush provided.
To obtain optimal accuracy when testing outdoors (sunlight), use the Cell
Cover when zeroing and reading the sample.
Meter is not compatible for use with powder pillows, tablets, and liquids from
other manufacturers.
Dip strip for entire countdown.
Each eXact
Strip Micro is valid for ONLY ONE TEST. Discard strip after use.
®
Dry the outside of the meter before storage.
Remove batteries before storing for prolonged periods.
Store the meter and test materials out of direct sunlight and away from
chemical storage areas.
Minimize exposure of meter and test reagents to heat above 90°F (32°C).
When installing batteries, make sure that the O-ring is still attached to the
screw before tightening.
To pair the eXact iDip
photometer with your smart device, open the app
®
and connect through the app. You CANNOT connect the devices in the
settings section of your smart device.
All results must be saved from the results page to history in order to
permanently save to your smart device.
The eXact iDip
photometer may experience Bluetooth
disconnection when
®
®
submerged in water too deep. To avoid this, collect water sample in a small
container/cup and pour or pipette (contact us for ordering) the water into the
Sample Cell.
To ensure lab quality results, it is recommended to clean the cell with the
brush provided after every test. Use Distilled White Vinegar to clean the
cell after testing for Phosphate or Iron to remove reagent deposits from
the cell wall.