Thank you for your eXact iDip purchase! This guide will quickly ® walk you through the technical details of your new eXact iDip ® This guide includes initial set-up, downloading the app, test procedures, and tips . You are on your way to smart digital water...
PATENTED SMART PHOTOMETER SYSTEM ® Using eXact Micro reagents in combination with the ® app and photometer completes the eXact iDip Smart ® Photometer System . Each test will require the use of one or ® more of the testing methods outlined in this manual .
FOR BEST ACCURACY All tests that begin with 570 are only for use with the eXact iDip 570 photometer . ® Test procedures can vary from test to test . Visit exactidip.com exactidip570. to read full instructions and watch instructional video for each test .
15 cm and 1 meter INSTALL “AAA” BATTERIES (NOT INCLUDED) 1. Use a #4 Phillips head screwdriver to remove the screw from the base of your eXact iDip ® 2. Remove the base. 3. Install Four (4) new AAA batteries as illustrated inside your iDip’s battery compartment.
GETTING STARTED DOWNLOAD THE APP Using your Smart Device, download the eXact iDip ® app . The eXact iDip app is the brain for this system; ® download the latest update to ensure you are using the current version with up-to-date tests and features .
COMPATIBLE SMART DEVICES Updated SMART PHONE COMPATIBILITY 05/2019 APPLE SAMSUNG SONY MOTOROLA iPhone 4s AND UP Galaxy Ace Style Xperia E1 Moto E Desire 610 iPhone SE Galaxy Alpha Xperia M2 Moto G Desire 816 iPod touch 5th AND UP Galaxy Core II Xperia T2 Ultra Moto X...
Replace batteries eXact iDip ® “Er1” on No result sent to Close & restart app . eXact iDip eXact iDip from app Reconnect to eXact iDip ® ® ® “Er2” on No connection Connect eXact iDip ® eXact iDip to app ®...
EXACT IDIP APP OVERVIEW ® NEWS: Icon will turn MENU: red to indicate Opens HOME SCREEN updates and slide-out for changes to easy access to the app all app features PROFILE: Changed HISTORY: preferred Units Accesses of Measure saved results...
.com/itssensafe . SETTINGS TIMEOUT The timeout setting is used to turn off the eXact iDip photometer . ® The default setting is 5 minutes, however this can be adjusted . To do this, tap either ‘+’...
USING THE EXACT IDIP ® HISTORY The History stores all your saved test result information and allows you to sort by date, customer name, or test type . HOW TO SORT BY CUSTOMER To sort by Customer begin by selecting ‘History’, then ‘Sort by’, ‘Customer’...
‘Buy’ (this will unlock the test in the app) . When purchasing tests in the app, be sure to select the appropriate test for your eXact iDip photometer (see page 7) .
USING THE EXACT IDIP ® PROFILE The Profile section of the app can be used to set your preferred units of measure for your tests . Another feature located in the Profile section is Archiving. To access the Profile section of the app, tap the person icon that appears in the top right of the screen throughout the app .
INSTRUCTIONS AND VIDEOS Test procedures can vary from test to test . Read full instructions and watch instructional video for each test as detailed below . TAP TEST NAME Tap the test name that appears in the test result window . A pop- up will display showing options for Instructions and Video .
. If not, ® select your eXact iDip from the bottom of the screen . ® Note : Ensure you always connect your eXact iDip photometer ® via the Bluetooth connection within the app. To verify connection ® look for the Bluetooth icon in the upper right corner of your photometer.
Tap ‘Select Test’ at the top (e .g . Alkalinity, Total) . The eXact iDip and app will both display the test being ® 570, all available tests run . If using the eXact iDip ® will begin with 570 (see below) . Alkalinity, Total AL Bluetooth Test...
SELECT, FILL, DIP, READ REMOVE STRIP Remove one eXact Strip Micro (e .g . Total Alkalinity) and ® set in a dry, convenient place . Replace cap on bottle . DIP STRIP AND PRESS READ Press to initiate a 20 second countdown and...
(vertical) and to the bottom of the CELL . Press READ . READ RESULT & RECORD Result is displayed on the eXact iDip and in the app . Record ® value . For additional tests, repeat steps 5-9 .
AUTO CALCULATIONS The eXact iDip app has the ability to automatically calculate results for various parameters once specific tests are completed . After completing the Required Tests below, go to the Results page of the app and the Desired Result values below will be displayed automatically .
Make sure the eXact iDip Apple Watch app is installed by opening the Watch app on the iPhone and scrolling down until you see eXact iDip . Tap eXact iDip and turn on “Show App on Apple Watch” . To use the Apple Watch with the...
MANAGING DATA RESULTS After tests have been performed, tap ‘Results’ at the bottom of the screen . ADD SITES Each set of results can be saved to a customers ‘Site’ (water source at the location). Select a site from the list or to add new sites, tap ‘Sites’, then ‘+’.
MANAGING DATA SAVE TO HISTORY In ‘Results’ screen, tap ‘Save’ to store into ‘History’ . If this step is omitted, test results will not be permanently saved . A ‘Saving Results’ pop-up appears, verifying that your result is now successfully saved . SEND/SHARE VIA EMAIL In History you can edit, select, and email your results .
MANAGING DATA SEND/SHARE VIA SOCIAL MEDIA To share your results on social media, use the ‘Select’ button and check the result(s) you want to share . Tap ‘Post’ at the top and choose whether to share via Facebook or Twitter . ARCHIVE page 14 for details on using the Archiving feature .
Bluetooth SMART is a low-power networking standard ® which uses short radio waves to allow electronic devices to communicate with each other wirelessly . The eXact iDip ® comes standard with Bluetooth 4 .0 technology . It is a class ®...
® same wavelength and delivers the same chemicals in the same proportions as AWWA method 3500-Cr B . The eXact iDip ® photometer cannot be used for compliance monitoring because the photometer wavelength of 570nm is outside of the required 490nm–530nm range .
USEPA, DIN and ISO reference test methods . The eXact iDip ® photometer has been factory calibrated and will stay valid because of its exceptional quality. We are so confident in the eXact iDip ® photometer, we offer an industry leading 2-year warranty .
METHOD VERIFICATION Ready Snap is a method verification solution ® with predetermined values to verify the accuracy of any manufacturer's tests including our eXact iDip Smart Photometer System . ® The easy 3 step procedure (snap, fill, and test) allows for quick verification of test parameters.
MODEL AQUATIC HEALTH CODE AND NSF/ANSI 50 CERTIFICATION The Model Aquatic Health Code (MAHC) is a set of guidelines published by the Centers for Disease Control and Prevention (CDC) . This document brings together the latest knowledge based on science and best practices to help state and local government officials develop and update pool codes.
EXACT IDIP TESTS & REAGENTS ® RANGE 1 % BEST # OF † PARAMETER / TEST PART # TESTS ACCURACY Alkalinity, Total 486641 40 – 200 7 .5 Alkalinity, Total Range Extender 486665 Each strip adds 130 ppm Bromine, Total (DPD–4) 486644 0 .00 –...