Buffers for calibration should be based on the measurement range.
Follow the calibration procedure properly.
Be sure to remove air bubbles from pH glass bulb for accurate measurements.
Store the meter and all test materials out of direct sunlight, heat, and away from
chemical storage areas.
Minimize exposure of meter and calibration solutions to heat above 100°F (38°C).
Dry the outside of the meter when testing is complete or before storage of the meter.
Pour pH 4.0 buffer solution up to the "FILL" line of the storage cap. Always store
the pH probe in clean 4.00 pH buffer solution.
Do not store the pH probe in a distilled or deionized water. This will cause
permanent damage to the probe.
Always fi ll the sample collection bottle with enough sample to completely
immerse the probe.
Rinse pH probe properly before putting in storage solution.
The meter has an automatic shut-off after 8 minutes of idle time.
This meter performs best in pool and typical tap water and does not give
reliable readings in distilled or deionized water.
Remove batteries when meter is not used for more than a month (Warranty Requirement).
Do not touch or rub the pH glass bulb.
To avoid cross-contamination, rinse the probe properly with clean water in
between measurements.
To clean the ORP sensor, dip the probe into 0.1 mol/L dilute hydrochloric
acid for 30 minutes; clean the platinum surface with dishwashing liquid; or use
plain white toothpaste to gently clean the platinum surface. Then, rinse with
clean water and soak for 6 hours in the KCl storage solution.

SAFETY INFORMATION

The manufacturer is not responsible for any damages due to
misapplication of this product including, but not limited to, direct,
indirect, incidental and consequential damages, and disclaims such
damages to the full extent permitted under applicable law. The user
is solely responsible to identify critical application risks and install
appropriate safeguards to protect processes during a possible
equipment malfunction. Caution: The pH probe has a glass bulb, do
not immerse directly in pool.

PROBE REPLACEMENT

1. Turn off the meter.
2. Unscrew the probe collar by turning counter-clockwise.
3. Gently pull downward to disconnect the probe from the meter.
Carefully plug in the replacement probe (pay attention to the probe
orientation when plugging into the meter).
4. Put back the probe collar and tighten fi rmly to secure the probe to
the meter (do not overtighten).