Electrical Requirements - Thermo Scientific Precision GP 02 User Manual

Precision series general purpose water baths
Table of Contents

Advertisement

Available languages
  • EN

Available languages

  • ENGLISH, page 67
Chapter 3
|
Installation
DANGER
CAUTION
CAUTION
3-2
|
Precision

Electrical Requirements

Refer to the bath's nameplate for specific electrical requirements.
The bath is intended for use on a dedicated outlet.
The bath construction provides protection against the risk of electrical shock by
grounding appropriate metal parts. The protection will not function unless the power
cord is connected to a properly grounded outlet. It is the user's responsibility to
assure a proper ground connection is provided.
The circuit protector on the rear of the bath is designed to protect the bath's internal components.
Note If the circuit protector activates allow the temperature to cool before resetting. Restart the
bath. Contact us if the circuit protector activates again.
The bath's electrical power cord is the disconnecting device, it must be easily
accessible at all times.
Ensure the cord does not come in contact with the reservoir contents.
Note Before inserting the electrical cord into the bath connection ensure the circuit protector is in
the O (off) position.
Once the cord is connected to the bath, connect the other end to the main power source.
The following plug types are available:
Bath
Volts
/Hertz/Phase
1
GP (All)
100-115/50-60/1
200-230/50-60/1
1. Volts ± 10%
Plug Type
N5-15
Country Specific
15 Amp Outlet
Thermo Scientific

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents