Setup

1.
Set up the 1488 HD console according to the instructions provided in
Stryker user manual P18966 or P18972.
2.
Connect the Pendulum Camera to the console.
• Unscrew the soaking cap from the cable connector if necessary.
• Align the blue arrow on the cable connector with the blue arrow
on the camera-connector port on the front console panel.
• Push in the connector until it locks in place.
• (To unplug the Pendulum Camera from the console, grasp the
knobbed portion of the connector and pull straight out.)
3.
Attach an endoscope to the Pendulum Camera.
• Remove the red dust cap if it is present.
• Lock the endobody brake by pushing it to the left.
• Twist the endobody clamp and hold it open.
5