Stryker 1080 Operation Manual
Stryker 1080 Operation Manual

Stryker 1080 Operation Manual

Fluoroscopy stretcher
Hide thumbs Also See for 1080:

Advertisement

Quick Links

Fluoroscopy Stretcher
1080
Operations Manual
2017/07 A.1
1080-109-001 REV A
www.stryker.com

Advertisement

Table of Contents
loading

Summary of Contents for Stryker 1080

  • Page 1 Fluoroscopy Stretcher 1080 Operations Manual 2017/07 A.1 1080-109-001 REV A www.stryker.com...
  • Page 2 sample text...
  • Page 3 General warning Caution Warning; crushing of hands No pushing Do not store the oxygen bottle Catalogue number Serial number Batch code For US Patents see www.stryker.com/patents CE mark EC REP Manufacturer Date of manufacture Safe working load Lubricate www.stryker.com 1080-109-001 REV A...
  • Page 4 sample text...
  • Page 5: Table Of Contents

    Disinfecting the mattress ........................18 Preventive maintenance ........................... 20 Warranty ............................... 21 Warranty exclusion and damage limitations .................... 21 To obtain parts and service ......................... 21 Return authorization........................... 21 Damaged product..........................21 International warranty clause ....................... 21 www.stryker.com 1080-109-001 REV A...
  • Page 6: Warning/Caution/Note Definition

    Note: Provides special information to make maintenance easier or important instructions clearer. 1080-109-001 REV A www.stryker.com...
  • Page 7: Summary Of Safety Precautions

    Improper usage of the product can cause injury to the patient or operator. Operate the product only as described in this manual. • Do not modify the product or any components of the product. Modifying the product can cause unpredictable operation resulting in injury to patient or operator. Modifying the product also voids its warranty. www.stryker.com 1080-109-001 REV A...
  • Page 8: Introduction

    If you have any questions, contact Stryker Customer Service or Technical Support at 1-800-327-0770. Product description The Stryker Model 1080 Fluoroscopy Stretcher is a wheeled stretcher that consists of a platform mounted on a wheeled frame to transport patients in a horizontal position within a healthcare facility by health professionals or trained representatives of the user facility.
  • Page 9: Specifications

    33 cm x 139.5 cm 27.5 in. x 66 in. 69.9 cm x 167.6 cm Fluoroscopy window 6 in. 15 cm Caster diameter/width 54 in. 137.2 cm C-arm access width Stryker reserves the right to change specifications without notice. www.stryker.com 1080-109-001 REV A...
  • Page 10: Product Illustration

    Portage, MI 49002 To view your operations or maintenance manual online, see https://techweb.stryker.com/. Have the serial number (A) of your Stryker product available when calling Stryker Customer Service or Technical Support. Include the serial number in all written communication. 1080-109-001 REV A...
  • Page 11: Serial Number Location

    Introduction Serial number location Date of manufacture The year of manufacture is the first 2 digits of the serial number. www.stryker.com 1080-109-001 REV A...
  • Page 12: Operation

    To lower the head end, push down on the hydraulic release pedal (B) at either end of the product. To lower the foot end, push down on the hydraulic release pedal (A) at either end of the product. To lower the entire litter, push down on both hydraulic release pedals (A and B) together. 1080-109-001 REV A www.stryker.com...
  • Page 13: Positioning The Product In Trendelenburg/Reverse Trendelenburg

    Note: Do not use siderails as restraint devices to keep the patient from exiting the product. The siderails keep the patient from rolling off the product. The operator must determine the degree of restraint necessary to make sure that the patient is safe. www.stryker.com 1080-109-001 REV A...
  • Page 14: Raising Or Lowering The Pneumatic Fowler Backrest

    To lower the Fowler backrest, squeeze the red Fowler backrest release handle. Push the Fowler backrest down to the desired position. Note: You can operate the Fowler backrest from both ends of the product litter. This allows the patient to be reversed on the product when taking X-rays. 1080-109-001 REV A www.stryker.com...
  • Page 15: Accessories

    Accessories These accessories may be available for use with your product. Confirm availability for your configuration or region. Call Stryker Customer Service: 1-800-327-0770. Name Part number Footboard/chart holder 1080-134-000 IV pole, 2-stage permanent, foot end 1080-112-000 IV pole, 2-stage permanent, head end...
  • Page 16: Positioning The Three-Stage Permanently Attached Iv Pole Option

    Do not hang IV bags that exceed 9.3 lb (4.2 kg) on a single IV hanger. • Always make sure that the IV pole is at a low height to allow the pole to pass safely through door openings when you transport a patient. 1080-109-001 REV A www.stryker.com...
  • Page 17: Attaching And Positioning The Removable Iv Pole

    Do not hang IV bags that exceed 40 lb (18 kg) on the IV pole. • Always make sure that the IV pole is at a low height to pass through door openings when you transport a patient. www.stryker.com 1080-109-001 REV A...
  • Page 18: Attaching The Upright Oxygen Bottle Holder

    Insert the support bar into any of the IV sockets. Insert the cotter pin through the hole in the support bar to secure the bottle holder to the product. Note: Do not use the upright oxygen bottle holder as a push/pull device. 1080-109-001 REV A www.stryker.com...
  • Page 19: Locating The Patient Restraint Strap Tie-Ins

    There are eight patient restraint strap tie-in locations on the litter assembly for attaching patient restraint straps (Figure 6 on page 15). Figure 6: Restraint strap tie-in locations www.stryker.com 1080-109-001 REV A...
  • Page 20: Cleaning

    Use a clean, soft cloth to wipe down the entire mattress with a mild soap and water solution to remove foreign material. Wipe down the mattress with a clean, dry cloth to remove any excess liquid or cleaning agents. 1080-109-001 REV A www.stryker.com...
  • Page 21: Remove Iodine

    Use standard household cleansers or vinyl cleansers and a soft bristle brush on troublesome spots or stains. Pre- soak dried-on soil. Laundering Laundering is not recommended. Laundering may substantially decrease the useful life of the mattress. www.stryker.com 1080-109-001 REV A...
  • Page 22: Disinfecting

    Phenolic disinfectant (active ingredient - o-phenylphenol) • Chlorinated bleach solution (5.25% bleach diluted 1 part bleach to 100 parts water which equals 520 ppm available chlorine (40 mL of a 5.25% bleach solution per 4000 mL water)) 1080-109-001 REV A www.stryker.com...
  • Page 23 • Frequent or prolonged exposure to higher concentrations of disinfectant solutions may prematurely age the cover fabric. • The use of accelerated hydrogen peroxides or quaternaries that contain glycol ethers may damage the cover. www.stryker.com 1080-109-001 REV A...
  • Page 24: Preventive Maintenance

    Remove product from service before you perform the preventive maintenance inspection. Check all items listed during annual preventive maintenance for all Stryker Medical products. You may need to perform preventive maintenance checks more often based on your level of product usage. Service only by qualified personnel.
  • Page 25: Warranty

    Stryker’s obligation under this warranty is expressly limited to supplying replacement parts and labor for, or replacing at its option, any product which is, in the sole discretion of Stryker, found to be defective. If requested by Stryker, product or parts for which a warranty claim is made shall be returned prepaid to the factory. Any improper use or any alteration or repair by others in such a manner as in Stryker’s judgment affects the product materially and...
  • Page 26 Stryker Medical 3800 E. Centre Avenue Portage, MI 49002 2017/07 1080-109-001 REV A www.stryker.com...