Specifications - AEMC SR661 User Manual

Ac current oscilloscope probe
Table of Contents

Advertisement

3.1
Electrical Specifications
*Reference conditions: 23°C ± 3°K, 20 to 75% RH, 48 to 65Hz, external magnetic field
<40A/m, no DC component, no external current carrying conductor, test sample centered.
Operating Range: 0.1 to 2000A peak
Measurement Range: 100mA to 1200Arms (2000A peak)
Output Signal: mV output signal (2V peak max)
100mV/A: 10mA to 20A peak
10mV/A: 0.1A to 200A peak
1mV/A:
1A to 2000A peak
1mV/A, (1V at 1000A)
Range
1 to 50A
50 to 200A
200 to 1000A
1000 to 1200A
Overload: 1200 A for 40 mn ON, 20 mn OFF
10 mV/A, (1V at 100A)
Range
0.1 to 5A
5 to 20A
20 to 100A
100 to 120A
Overload: 120 A continuous
100mV/A, (1V at 10 Arms)
Range
0.1 to 0.5A
0.5 to 2A
2 to 10A
10 to 12A
Overload: 12 A continuous
AC Current Oscilloscope Probe Model SR661
CHAPTER 3

SPECIFICATIONS

Accuracy
≤1% ± 1mV
Accuracy
≤2% ± 5mV
Accuracy
≤3% ± 10mV
Phase Shift
N/A
≤3°
≤2°
≤1°
Phase Shift
N/A
≤15°
≤10°
≤5°
Phase Shift
N/A
N/A
≤15°
7