AEMC SL261 User Manual

Ac/dc current oscilloscope probe
Hide thumbs Also See for SL261:

Advertisement

Quick Links

SL261
AC/DC CURRENT
OSCILLOSCOPE PROBE
E N G L I S H
User Manual

Advertisement

Table of Contents
loading

Summary of Contents for AEMC SL261

  • Page 1 SL261 AC/DC CURRENT  OSCILLOSCOPE PROBE E N G L I S H User Manual...
  • Page 2 The recommended calibration interval for this instrument is 12 months and begins on the date of receipt by the customer. For recalibration, please use our calibration services. Refer to our repair and calibration section at www.aemc.com. Serial #: ________________________________ Catalog #: 1201.51 Model #:...
  • Page 3: Table Of Contents

    Table of Contents 1. INTRODUCTION ................2 1.1 International Electrical Symbols ..........3 1.2 Definition of Measurement Categories ........3 1.3 Receiving Your Shipment ............3 1.4 Ordering Information ..............4 1.4.1 Accessories and Replacement Parts ......4 2. PRODUCT FEATURES ..............5 2.1 Description ................5 2.2 Compatibility ................5 2.3 Features ..................6 3. SPECIFICATIONS................7 3.1 Electrical Specifications ............7 3.2 Mechanical Specifications ............8 3.3 Environmental Specifications ............9...
  • Page 4: Introduction

    Read the instruction completely. • Use caution on any circuit: potentially high voltages and currents may be present and may pose a shock hazard. • Do not use the probe if damaged. Always connect the current probe to the measuring device before it is connected around the conductor • Do not use on non insulated conductor with a potential to ground greater than 600V CAT III pollution 2. Use extreme caution when clamping around bare conductors or bus bars. • Before each use, inspect the probe; look for cracks in housing or output cable insulation. • Do not use clamp in wet environment or in locations that hazard- ous gases exist. • Do not use the probe anywhere beyond the tactile barrier. AC/DC Current Oscilloscope Probe Model SL261...
  • Page 5: International Electrical Symbols

    CAT IV: For measurements performed at the primary electrical supply (<1000V) such as on primary overcurrent protection devices, ripple control units, or meters. Receiving Your Shipment Upon receiving your shipment, make sure that the contents are consistent with the packing list. Notify your distributor of any missing items. If the equipment appears to be damaged, file a claim immediately with the car- rier and notify your distributor at once, giving a detailed description of any damage. Save the damaged packing container to substantiate your claim. AC/DC Current Oscilloscope Probe Model SL261...
  • Page 6: Ordering Information

    Ordering Information AC/DC Current Probe Model SL261 (10A-100mV/A, 100A-10mV/A, BNC) ........Cat. #1201.51 Includes a user manual. 1.4.1 Accessories and Replacement Parts Adapter - BNC Adapter for use with AC/DC Current Probe Model SL261 & Meters 8220, 8333, 8335 & 8336 ....Cat. #2140.40 AC/DC Current Oscilloscope Probe Model SL261...
  • Page 7: Product Features

    CHAPTER 2 PRODUCT FEATURES Description The SL261 AC/DC Current Oscilloscope Probe expands oscilloscope applications in industrial, automotive or power environments, and is ideal for analysis and measurement of distorted current waveforms and har- monics. The probe permits accurate display and measurement of currents from 100mA to 100Arms, DC to 100kHz without breaking into the circuit. The probe uses Hall effect technology to measure AC and DC signals. The probe connects directly to an oscilloscope through a 2 meter coaxial cable with an insulated BNC. Compatibility The Model SL261 is compatible with any analog or digital oscilloscope or other voltage-measuring instrument which has the following features: • BNC input connector • Range capable of displaying 0.2 to 0.5V per division • Minimum input impedance of 1MΩ AC/DC Current Oscilloscope Probe Model SL261...
  • Page 8: Features

    Range Selection Switch Ø = 0.46" (11.8 mm) Keep hands below this point Peak Overload Indication LED Battery Compartment Battery Screw Compartment Cover WARNING: Always keep hand on the handle part of the probe. AC/DC Current Oscilloscope Probe Model SL261...
  • Page 9: Specifications

    ± 4% of reading ± 50mA (500mA to 40A peak) DC/AC 10mV/A 12% of reading ± 50mA < 1° from DC to 65Hz (40A to 80A peak) DC/AC 10mV/A 15% of reading (80A to 100A peak) DC/AC Frequency Range: DC to 100kHz (-3dB with current derating) Noise: Range 10mV/A: 480µV Range 100mV/A: 3mV Slew Rate: Range 10mV/A: 20mV/µs Range 100mV/A: 0.3V/µs Load Impedance: > 1MΩ/100pF AC/DC Current Oscilloscope Probe Model SL261...
  • Page 10: Mechanical Specifications

    Zero Adjustment: 20 turn potentiometer Maximum Cable Diameter: 0.46" (11.8 mm) Case Protection: IP20 per IEC 529 Drop Test: 1.0m on 38mm of oak on concrete; test according to IEC 1010 Mechanical Shock: 100G; test per IEC 68-2-27 Vibration: Test per IEC 68-2-6, frequency range 10 Hz to 55 Hz, amplitude 0.15mm Handle: Lexan 920A, UL 94 V2 ® Dimensions: 9.09 x 1.42 x 2.64" (231 x 36 x 67mm) Weight: 11.6 oz (330g) with battery Color: Light gray Output Lead: Insulated coaxial cable with insulated BNC connector Cable Length: 6.5 foot (2m) AC/DC Current Oscilloscope Probe Model SL261...
  • Page 11: Environmental Specifications

    Influence of Temperature: < 0.2% per °C Altitude: Non-operating: 0 to 12,000m Operating: 0 to 2000m Safety Specifications E192383 Electrical: Double insulation or reinforced insulation between the primary or secondary and the outer case of the handle per EN 61010-2-032. 600V, CAT III, Pollution Degree: 2 Electromagnetic Compatibility: EN 50081-1 Class B EN 50082-2 Electrostatic discharge IEC 1000-4-2 Radiated Field IEC 1000-4-3 Fast Transients IEC 1000-4-4 Magnetic Field at 50/60 Hz IEC 1000-4-8 *Specifications are subject to change without notice AC/DC Current Oscilloscope Probe Model SL261...
  • Page 12: Typical Response Curves

    10 mV/A frequency response measurement current: 1A sine Frequency 10000 phase in ° 100000 Frequency in Hz range 10 mV/A -100 phase graph measurement current: 1A sine -150 range 100 mV/A Phase Shift AC/DC Current Oscilloscope Probe Model SL261...
  • Page 13 Typical Specification -10% -12% -14% -16% 100A Accuracy AC/DC Current Oscilloscope Probe Model SL261...
  • Page 14: Operation

    1000mV peak or 2000mV peak to peak maximum. The peak ratings by range are: 10A peak on the 100mV/A range, 100A peak on the 10mV/A range. AC/DC Current Oscilloscope Probe Model SL261...
  • Page 15: Battery Indication (Green Led)

    Battery Indication (Green LED) The probe has a battery condition LED. To ensure proper readings with your current probe, be sure that the green LED is lit during measurement. If not, replace the 9V battery. Peak Overload (OL) Indication (Red LED) The SL261 offers an overload indicator. If the red LED illuminates during measurement, this indicates that the peak value exceeds the instrument response level and that the output is distorted. Switch the probe to a higher range if possible. AC/DC Current Oscilloscope Probe Model SL261...
  • Page 16: Maintenance

    • To avoid electrical shock and/or damage to the instrument, do not allow water or other foreign substances into the case. • Disconnect the unit from all circuits and test cables before open- ing the case. Cleaning • To clean the probe body, use a soft cloth dampened in a solution of mild detergent and water. To clean the core, open the jaw and clean the exposed core surfaces with a cotton swab dampened with isopropyl alcohol (isopropanol) or ethyl alcohol (fotocol or ethanol). Lubricate the jaws mating surfaces with a light oil. • Do not use chemicals containing benzine, benzene, toluene, xylene, acetone, or similar solvents. • Do not immerse the probe in liquids or use abrasive cleaners. AC/DC Current Oscilloscope Probe Model SL261...
  • Page 17: Replacing The Batteries

    If the green battery indication LED does not light up when the probe is turned on, the batteries will need to be replaced. Do not replace the battery while probe is in use. To replace the battery: • Disconnect the probe from the circuit and the oscilloscope. • Turn the probe “OFF”. • Unscrew the battery compartment screw and pull out the battery com- partment cover. • Replace the battery and put the cover back on. AC/DC Current Oscilloscope Probe Model SL261...
  • Page 18: Repair And Calibration

    NOTE: You must obtain a CSA# before returning any instrument. Technical and Sales Assistance If you are experiencing any technical problems, or require any assistance with the proper operation or application of your instrument, please call, mail, fax or e-mail our technical support team: Chauvin Arnoux , Inc. d.b.a. AEMC Instruments ® ® 200 Foxborough Boulevard Foxborough, MA 02035 USA Phone: (800) 343-1391 (508) 698-2115 Fax: (508) 698-2118 E-mail: techsupport@aemc.com www.aemc.com NOTE: Do not ship Instruments to our Foxborough, MA address. AC/DC Current Oscilloscope Probe Model SL261...
  • Page 19: Limited Warranty

    Limited Warranty The AC/DC Current Oscilloscope Probe Model SL261 is warranted to the owner for a period of two years from the date of original purchase against defects in manufacture. This limited warranty is given by AEMC Instruments, ® not by the distributor from whom it was purchased. This warranty is void if the unit has been tampered with, abused or if the defect is related to service not performed by AEMC Instruments. ® Full warranty coverage and product registration is available on our website at www.aemc.com/warranty.html.
  • Page 20 04/17 99-MAN 100042 v16 Chauvin Arnoux , Inc. d.b.a. AEMC Instruments ® ® 15 Faraday Drive • Dover, NH 03820 USA • Phone: (603) 749-6434 • Fax: (603) 742-2346 www.aemc.com...