AC Current Probe
Model SR634
User Manual
DESCRIPTION
The SR634 (Catalog #2113.48) is designed for use in industrial environments. The ergonomic design
allows it to easily attach to cables or small bus bars. The "circular" jaws guarantee a very good accuracy
and low phase shift. The probe has a measurement range up to 1000Arms. continuous and is compatible
with any AC Ammeter, multimeter, or other current measurement instrument with an input impedance lower
than 0.6Ω. To achieve the stated accuracy, use the SR634 with an ammeter having an accuracy of 0.75%
or better.
WARNING
The safety warnings are provided to ensure the safety of personnel and proper operation of the
instrument. Read the instruction completely.
• Use caution on any circuit: potentially high voltages and currents may be present and may pose a
shock hazard.
• Do not use the probe if damaged. Always connect the current probe to the measuring device before
it is connected around the conductor
• Do not use on non-insulated conductor with a potential to ground greater than 600V CAT III pollution
2. Use extreme caution when clamping around bare conductors or bus bars.
• Before each use, inspect the probe; look for cracks in housing or output cable insulation.
• Do not use clamp in wet environment or in locations that hazardous gases exist.
• Do not use the probe anywhere beyond the tactile barrier.
INTERNATIONAL ELECTRICAL SYMBOLS
This symbol signifies that the current probe is protected by double or reinforced insulation. Use
only factory specified replacement parts when servicing the instrument.
This symbol signifies CAUTION! and requests that the user refer to the user manual before using
the instrument.
This is a type A current sensor. This symbol signifies that application around and removal from
HAZARDOUS LIVE conductors is permitted.
DEFINITION OF MEASUREMENT CATEGORIES
CAT II: For measurements performed on circuits directly connected to the electrical distribution system.
Examples are measurements on household appliances or portable tools.
CAT III: For measurements performed in the building installation at the distribution level such as on
hardwired equipment in fixed installation and circuit breakers.
CAT IV: For measurements performed at the primary electrical supply (<1000V) such as on primary
overcurrent protection devices, ripple control units, or meters.
RECEIVING YOUR SHIPMENT
Upon receiving your shipment, make sure that the contents are consistent with the packing list. Notify your
distributor of any missing items. If the equipment appears to be damaged, file a claim immediately with the
carrier and notify your distributor at once, giving a detailed description of any damage.
99-MAN 100134.v10 03/17

Advertisement

loading

Summary of Contents for AEMC SR634

  • Page 1 The probe has a measurement range up to 1000Arms. continuous and is compatible with any AC Ammeter, multimeter, or other current measurement instrument with an input impedance lower than 0.6Ω. To achieve the stated accuracy, use the SR634 with an ammeter having an accuracy of 0.75% or better.
  • Page 2: Electrical Specifications

    Range 500:5 - 10mA/A ELECTRICAL SPECIFICATIONS Primary 1 to 10A 25A 100A 500A 600A Current Range: ≤ 6% ± 0.1A Accuracy , continuous @ ≤ 1kHz 1 to 1000A Phase 6° 4° 3° 2.5° 2.5° Overload: 600A for 30 mn on, 15 mn OFF Transformation Ratio: 250:5, 500:5, 1000:5 Range 1000:5 - 5mA/A Output Signal: 5mA, 10mA, 20mA AC/A AC...
  • Page 3: Safety Specifications

    Range 500:5 - < 0.4% & < 0.6° Range 1000:5 - < 0.2% & < 0.2° Jaw Opening: ORDERING INFORMATION 2.25" (57 mm) max. AC Current Probe SR634 ....Cat #2113.48 Maximum Conductor Size: Accessories: 2.05" (52 mm) Banana plug adapter Envelope Protection: (to non-recessed plug) ....
  • Page 4 Making Measurements with the AC Current Probe Model SR634 • Connect the current probe to an AC current range on your DMM or other current measuring instrument. The AC Current Probe SR634 is a multi-ratio unit. The outputs are as follows: Current Flow Output of Probe...