Messages and Diagnostics
Error
Data and
No.
command errors -
description
parameter
76
Value range
Adjustment digits
exceeded
77
Regulation code not
permitted
78
Zero setting range
> 4% or zero setting
range exceeded
79
Number of weight
ranges not allowed
80
81
Relationship
between weighing
ranges implausible
82
Numeral step not
plausible
83
Filter parameter
implausible
84
Characteristic value
not plausible
85
Averaging filter
length not permitted
86
Adjustment weight
error
87
88
7-144
Solution
The adjustment digits definition is too large. Enter a smaller
value. The highest permitted value is 16 777 215.
The regulation code is incorrect. Enter the proper code.
Permitted is ---- or
OIML..
The zero definition range or zero setting range is too large. This
is the case if you have entered a zero setting range > 4% of the
maximum weighing range or > 20 % in the zero setting range in
DR 3 in calibratable operation (Specifications Code OIML in DR
3). The sum of negative and positive values always apply here.
Reduce the zero definition range or zero setting range.
The number of the weighing range is incorrect. Enter 1, 2 or 3.
Reserved
The definitions for the weighing range (Max. and Numerical step)
are incorrect.
With a multi-segment scale, the min-value of the next higher
range must equal the max-value of the next lower range. The
max-value must be higher than the min-value.
With a multi-range scale, the min-value and max-value in
sequential ranges must always be greater and the max-value of
a range just always be greater than the min-value.
Check parameter definitions.
At least one numeral step in DR 3 is incorrect.
The following numeral steps are permitted as numeral steps for
the three weighing ranges and the distribution value: 50, 20, 10,
5, 2, 1, .., 0.5, 0.2, 0.1, 0.05, 0.02, 0.01, 0.005, 0.002, 0.001,
0.0005, 0.0002, 0.0001.
In calibratable operation (Specifications Code OIML in DR 3) the
numeral step in all three weighing ranges may not be less than
one 6000th of the weighing range max-value.
In calibratable operation, the total numeral step (Specifications
Code OIML in DR 3) may not be less than the numeral step of
weighing range 1.
The numeral step of a weighing range may not be less than the
numeral step in the next lower lying weighing range.
Check the parameter.
Filter parameter definitions in DR 3 not plausible.
Check the codes for the filter type (0..2) and the limit frequency
(0..9).
The characteristic value defined in DR 3 is incorrect. Set the
parameter to a valid value (1, 2 or 4).
The mean value filter in DR 3 has incorrectly defined parameters.
Set the parameter for the depth of the mean value filter to
maximum 250.
The defined adjustment weights are incorrect. The weight values
must increase or be 0 if not used.
Reserved
Reserved
SIWAREX FTA