Weighing Functions
Name
Dosing filter type
BYTE
Dosing filter limit
BYTE
frequency
Tare / Zero setting
Tare / Zero setting mode
BYTE
Tare / Zero setting cycle
BYTE
Reserve 3
WORD
Tare min. weight
REAL
Tare max. weight
REAL
Cycle time for zero
TIME
setting
Step control / check stop
Step control through
BYTE
digital input 1
Step control through
BYTE
digital input 2
5-96
Type
Address
Default
DBB022
0
DBB023
2
0
DBB024
0
DBB025
0
DBB026
0
DBB028
0
DBB032
0
DBB036
300000
msec
DBB040
0
DBB041
0
Range of Values / Explanation
Filter type for dosage control
0: critically damped
1: Bessel-Filter
2: Butterworth Filter
Other definitions not permitted.
0: No filter
1: fg = 20Hz
2: fg = 10Hz
3: fg = 5Hz
4: fg = 2Hz
5: fg = 1Hz
6: fg = 0,5Hz
7: fg = 0,2Hz
8: fg = 0.1Hz
9: fg = 0´.05Hz
Other definitions not permitted.
0: No taring, no zero setting at weighing start
1: Zero setting
2: Tare
3: Taring with average value
4: Taring using external entry tare
Other definitions not permitted.
0: Every weighing is nullified or tared
1: Tipping is not nullified or tared
2...99: 2...99 tippings not nullified or tared
Other definitions not permitted.
Reserve 3
Taring or ext. tare entries are only executed if
gross > tare minimum weight
0: No monitoring of the tare minimum weight
Tarring or ext. Tare entries are only executed if
gross < tare maximum weight
0: No monitoring of the tare maximum weight
If 0, then no time controlled zero setting
Unequal 0: Time between two zero settings
Note:
Zero setting / taring is performed after 15min at
the latest with weighing operating mode AWI and
country code "OIML"
Instead of defined command codes for the digital
inputs, step-on conditions can be influenced
through the inputs for dosing control (prerequisite
in DR interface parameters is identification 0xFF).
0: Weighing waits at step 0 if DI1 is active;
1: Weighing waits at step 1 if DI1 is active
2: Weighing waits at step 2 if DI1 is active
...
7: Weighing waits at step 7 if DI1 is active
Other definitions not permitted.
Instead of defined command codes for the digital
inputs, step-on conditions can be influenced
through the inputs for weighing control
(prerequisite in DR interface parameters is
identification 0xFF).
0: Weighing waits at step 0 if DI2 is active;
1: Weighing waits at step 1 if DI2 is active
2: Weighing waits at step 2 if DI2 is active
...
7: Weighing waits at step 7 if DI2 is active
Other definitions not permitted.
Reference
5.16.8
5.16.10
5.16.11
5.16.12
5.16.13
5.16.14
5.16.15
5.16.16
5.16.16
SIWAREX FTA