Patient Transport - Arjohuntleigh citadel Instructions For Use Manual

Patient therapy system
Hide thumbs Also See for citadel:
Table of Contents

Advertisement

two seconds to turn off the control unit. When the unit turns off, you
will hear the valves open and the mattress will deflate.

Patient Transport

1. Press the Patient Transport Button to inflate the Air Mattress slightly
above the set pressures to prepare for patient transport.
2. You may unplug the power cord from the wall outlet and wrap the
cord onto the cord hook at the head end of the unit once the bed has
emitted an audible tone and the transport indicator is solid green.
3. If necessary, place patient's IV therapy onto IV poles that may be
placed into sockets located on all four corners of the unit.
4. Verify side rails are raised and locked.
5. Unlock brakes.
6. Transport patient following all applicable safety rules and institution
protocols.
7. Plug power cord into wall outlet immediately following completion of
patient transport.
8. Once bed is plugged into the wall outlet, press and hold the On / Off
button on the mattress control unit for two seconds to turn on the
control unit.
Citadel Patient Therapy System
25

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Citadel c200Citadel c100

Table of Contents