Arjohuntleigh citadel Instructions For Use Manual page 27

Patient therapy system
Hide thumbs Also See for citadel:
Table of Contents

Advertisement

solid amber
solid amber
solid amber
solid amber
solid amber
solid amber
Citadel Patient Therapy System
Active Alarm Icons
solid amber
solid amber
flashing green
flashing green
flashing green
flashing green
solid amber
Alarm Description
Head angle raised above
30° during patient turn
(nurse assist).
flashing green
Head angle raised above
30° during continuous
patient turn.
flashing green
Mattress cannot reach
target pressure after
10 minutes. Tighten air
connectors See page 15.
Mattress base cannot
reach pressure after
five minutes. Tighten air
connectors See page 15.
Turn bladder cannot
reach pressure in patient
turn (nurse assist) after
five minutes. Tighten air
connectors See page 15.
Turn bladder cannot reach
pressure in continuous
patient turn after five
minutes. Tighten air
connectors See page 15.
CPR switch cable
disconnected. Reconnect
CPR switch cable.