EMC INFORMATION

When this equipment is used adjacent to other electronic
devices, the user should observe the equipment to verify normal
operation.
Guidance and manufacturer's declaration - electromagnetic emissions
The Citadel Mattress Control Unit is intended for use in the electromagnetic environment
specified below. The customer or the user of the Citadel Mattress Control Unit should ensure
that it is used in such an environment.
Emissions test
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic
emissions
IEC 61000-3-2
Voltage
fluctuations/
flicker
emissions
IEC 61000-3-3
Guidance and manufacturer's declaration – electromagnetic immunity
The Citadel Mattress Control Unit is intended for use in the electromagnetic environment
specified below. The customer or the user of the Citadel Mattress Control Unit should ensure
that it is used in such an environment.
Immunity test
Electrostatic
discharge
(ESD) IEC
61000-4-2
Electrical fast
transient/burst
IEC 61000-4-4
Citadel Patient Therapy System
Compliance
Group 1
Class A
Class A
Complies
IEC 60601
test level
± 6kV contact
± 8kV air
± 2kV for power
supply lines
± 1kV for input/
output lines
Electromagnetic environment - guidance
The Citadel Mattress Control Unit uses RF energy only
for its internal function. Therefore its RF emissions are
very low and are not likely to cause any interference in
nearby electronic equipment.
The Citadel Mattress Control Unit is suitable for use
in all establishments other than domestic and those
directly connected to the public low-voltage power
supply network that supplies buildings used for
domestic purposes.
Compliance level
± 6kV contact
± 8kV air
± 2kV
Not applicable
Electromagnetic environment
- guidance
Floors should be wood,
concrete or ceramic tile. If floors
are covered with synthetic
material, the relative humidity
should be at least 30%.
Mains power quality should be
that of a typical commercial or
hospital environment.