Setting The Agitation Time - IKA HS 260 Basic Operating Instructions Manual

Orbital shakers
Hide thumbs Also See for HS 260 Basic:
Table of Contents

Advertisement

Available languages

Available languages

Note: If the time expiration signal has been activated, expiration of the
set time is indicated by a sequence of signals (3 times 5 short tons at
intervals of about 10 seconds).
Adjusting the lock function
The lock function of the agitation table can be activated or deac-
tivated with a direct connection to the setting of the audio time
expiration signal.
The following steps are required to activate or deactivate the
function:
a.)
Turn off the device with the
power switch.
b.)
Hold down the right turn
button and turn on the devi-
ce with the power switch.
Let go of the turn button
after about 5 seconds. After
you have run through the
Sound On/Off routine, it is
possible to activate or deac-
tivate the lock function.
c.)
You can activate (Lock On) or
deactivate (Lock Off) the lock-
function by pressing the right
turnbutton.
The activated lock function is displayed by the
right display. If you cannot see the = icon in the right display,
the lock function of the agitation table is turned off.
26
ATTENTION!
(freely movable) agitation table will continue to perform
a circular motion until the lock function can fasten the
agitation table in place.

Setting the agitation time

Basic model
The amount of time required for the agitation process can be set with
the aid of the right turn button. A distinction is made for the agitation
time setting between timed mode and continuous operation. If conti-
nuous mode is selected (On LED is lit), the agitator can continue its agi-
tation function for any amount of time with the previously set speed.
If timed mode is selected, the agitation process can last for a maximum
of 50 min. The set time is displayed by the LED line. The flashing LED
shows the amount of time currently remaining After all LED's have
gone out in timed mode, the device switches off into the Off operating
mode and the agitation process comes to an end.
Note: The curently set time can be modified at any time. If the time
setting is set to the right beyond 50 minutes, the LED bar goes out
and the LED in the On operating state (continuous mode) starts to
light up. If you continue to hold down the turn button, the device first
jumps to the Off operating mode (standstill) and then back into timed
operation. The same applies in the opposite order if the turn button
=
icon in the
is held down beyond a certain amount of time and moved to the left.
After the power failed, the set time (timed mode or continuous
mode) is erased and the device switches into Offline operating mode
(standstill).
After a power failure, the unlocked

Advertisement

Table of Contents
loading

This manual is also suitable for:

Ks 260 basicKs 260 controlHs 260 control

Table of Contents