Adjust the temperature limit (increase the value by clock-
wise / decrease the value by counterclockwise) with screw-
driver delivered with the device�
Check whether the voltage specified on the type plate
matches the mains voltage available�
The power socket used must be
earthed (protective earthing)�
If above conditions are met, the device is ready for opera-
tion after plugging in the mains plug�
If these conditions are not met, safe operation is not guar-
anteed and the device could be damaged�
Observe the ambient conditions (temperature, humidity,
etc�) listed under "Technical data"�
Switch on:
After switching on the mains switch (1), all LED segments
light up during the self test�
Starting
Device type
Temperature limit
After that, following two screens appear alternatively�
Room temperature
Then the device enters standby status and is ready for op-
eration�

Setting the temperature limit

+
-

Commissioning

Operating mode (A) Software version
The temperature limit setting will appear on the display�
Temperature limit
Factory setting: approximate max. value
Adjustment range: see "Technical data"�
The device may only be used
WARNING
for heating up non-explosive
and non-flammable substances
and objects.
Heating function:
The heating function can be started or stopped by pressing
the "Start/Stop" key (I)�
When the heating function is activated by pressing the "Start/
Stop" key (I), the heating indicator (T) lights up�
Then, the target temperature and the actual temperature
appears on display (A) alternately� Accordingly, setting value
indicator (S) or actual value indicator (R) lights up alternately�
The heating temperature can be adjusted by pressing the
"Temp (+)" key (O) or "Temp (-)" key (P)�
By pressing the "Temp (+)" key (O) or "Temp (-)" key (P), the
temperature increase or decrease in 0�1 °C steps�
For changing the temperature more quickly, Keep pressing
on "Temp (+)" key (O) or "Temp (-)" key (P)�
The value could be changed in standby or operation process�
Flap open function:
Change the position of the flap by pressing the falp open
+/- keys (in 10 % increments)�
Counter and timer function:
Counter function:
If the "T mode" is deactivated, when the heating function
is started by pressing "Start/Stop" key (I), the counter will
start automatically and counter indicator (D) lights up� The
counter value shows on display (C)�
When the operating time is less than 1 day (24 hours), the
counter work with hour/minute (hh:mm) mode�
When the operating time is more than 1 day (24 hours), the
display switches from hour/minute mode to day mode� In
the day mode, the display show the day and hour/minute
(hh:mm) alternatively�
When the heating function is stop by pressing "Start/Stop"
key (I), the counter value appear on display (C)�
When the heating function is started again, the counter will
re-start from 00:00�
Note: The maximum counter value is 100 days (d: 99/23:59)�
During counter mode, the "Timer" key is invalid�