Keithley 6430 Quick Results Manual page 38

Sub-femtoamp remote sourcemeter
Hide thumbs Also See for 6430:
Table of Contents

Advertisement

Remote command programming
SCPI commands
SCPI commands for linear, logarithmic and custom sweeps are listed in Table 16.
NOTE
Table 16 does not provide a complete listing of sweep commands. Documentation
for all sweep commands can be found in Section 9 of the Model 6430 SourceMeter
Instruction Manual.
Table 16
SCPI commands: sweeping
*
Command
:SOURce:xxx:MODE <name>
:SOURce:DELay <n>
:SOURce:DELay:AUTO <b>
:TRIGger:COUNt <n>
Linear and logarithmic sweeps:
:SOURce:SWEep:SPACing <name> Select sweep scale: <name> = LINear or
:SOURce:SWEep:RANGing
<name>
:SOURce:xxx:STARt <n>
:SOURce:xxx:STOP <n>
:SOURce:xxx:STEP <n>
:SOURce:SWEep:POINts <n>
Custom (list) sweep:
:SOURce:LIST:xxx <list>
:SOURce:LIST:xxx:APPend <list>
:SOURce:LIST:xxx:POINts?
Source memory sweep:
:SOURce:FUNCtion MEM
:SOURce:MEMory:SAVE <n>
:SOURce:MEMory:RECall <n>
:SOURce:MEMory:STARt <n>
:SOURce:MEMory:POINts <n>
*:xxx = :VOLTage or :CURRent.
Quick Results Guide
Description
Select source mode: <name> = FIXed, LIST or
SWEep.
Set source delay: <n> = 0 to 9999.9980 (sec).
Enable/disable auto delay: <b> = ON or OFF.
Set trigger count: <n> = Number of points in
sweep.
LOGarithmic.
Select source ranging: <name> = BEST, AUTO, or
FIXed.
Specify sweep start level: <n> = V or I source
value.
Specify sweep stop level: <n> = V or I source
value.
Specify linear sweep step level: <n> = V or I source
value.
Set number of sweep points: <n> = 2 to 2500.
Define source list: <list> = value1, value2, ...
valueN.
Add source list value(s): <list> = value1, value2, ...
valueN.
Query length of source list.
Select source memory sweep mode.
Save setup in memory location: <n> = 1 to 100.
Return to specified setup: <n> = 1 to 100.
Set source memory start location: <n> = 1 to 100.
Specify number of sweep points: <n> = 1 to 100.
31