Command Sequence For Auto Ohms Example; Command Sequence For Manual Ohms Example - Keithley 6430 Quick Results Manual

Sub-femtoamp remote sourcemeter
Hide thumbs Also See for 6430:
Table of Contents

Advertisement

18
Quick Results Guide
Auto ohms example — Table 11 shows a typical command sequence to use auto ohms to
measure resistance.
Manual ohms example — Table 12 shows a typical command sequence to use manual ohms
to measure resistance. This example uses auto range for current measurements. If you were to
use manual ranging, you should select the lowest possible current range for the measurement.
For example, to measure a 1M resistor with the source set to 2V, the 1µA measure range
should be selected to achieve the best accuracy (2V/1µA = 2M).
Table 11

Command sequence for auto ohms example

Command*
*RST
:SENS:FUNC "RES"
:SENS:RES:RANG 200
:SENS:RES:MODE AUTO
:OUTP ON
:READ?
*SourceMeter must be addressed to talk after sending :READ? to trigger and acquire data.
Table 12

Command sequence for manual ohms example

Command*
*RST
:SENS:FUNC "RES"
:SENS:RES:MODE MAN
:SOUR:FUNC VOLT
:SOUR:VOLT:MODE FIX
:SOUR:VOLT:RANG 2
:SOUR:VOLT:LEV 2
:SENS:CURR:PROT 10e-3
:SENS:CURR:RANG:AUTO
ON
:OUTP ON
:READ?
*SourceMeter must be addressed to talk after sending :READ? to trigger and acquire data.
Comments
Restore GPIB defaults (one-shot measurement mode).
Select ohms measurement function.
Select 200 range.
Select the auto ohms measurement method.
Turn the output on.
Trigger and acquire one data string.
Comments
Restore GPIB defaults (one-shot measurement mode).
Select ohms measurement function.
Select the manual ohms measurement method.
Select voltage source function.
Select fixed voltage sourcing mode.
Select 2V source range.
Set source level to 2V.
Set current compliance to 10mA.
Select auto range for current measurements.
Turn the output on.
Trigger and acquire one data string.