Mobile units, baskets and
inserts
To ensure that the mobile units, baskets
and inserts function properly, they must
be checked daily. A checklist is
supplied with the washer-disinfector.
Check the following points:
– Are the rollers in proper condition
and are they securely attached to the
mobile unit / insert?
– Are the counter-nuts on the guide
rails firmly secured?
– Is the mobile unit connection
correctly set and firmly screwed on?
– For mobile units in the modular
system, are the caps in the module
connections working properly?
– Are all spray jets, spray sleeves and
hose adapters securely attached to
the mobile unit / insert?
– Are all spray jets, spray sleeves and
hose adapters unclogged so that
water can flow through?
– Are all caps, covers and fasteners
securely attached to the spray
sleeves?
– Is the magnetic strip of the Automatic
Mobile Unit Recognition free of
adhering metallic items?
– Are the screws in the magnet rails for
Automatic Mobile Unit Recognition
tightly secured?
Maintenance of the washer-disinfector
If there are spray arms:
– Do the spray arms rotate freely?
– Are the spray jets clogged? See the
section on "Cleaning and Care /
Cleaning the spray arms".
– Are the magnets in place on the
spray arm ends?
– Are the magnets on the spray arms
free of attached metal parts?
Mobile units, modules and inserts
must also undergo routine
maintenance within the framework of
periodical maintenance of the
washer-disinfector every 2000
operating hours or at least once
per year.
See Maintenance of the
washer-disinfector.
67