The Model MR6292 probe is used in conjunction with the Micro-Ohmmeter
Model 6292 to perform measurements in BSG (both sides grounded) mode.
For additional information about using the Micro-Ohmmeter Model 6292,
refer to the user manual on the USB Stick that ships with the instrument.
4.1
1. Plug the probe into the Model 6292.
3. Turn the power to the Model 6292 ON.
4. The green LED on the Model MR6292 should be lit. If the LED does
not come ON or goes OFF after a few minutes; check to ensure that
the Model 6292 instrument is still powered ON, and that the Model
MR6292 is securely plugged into the instrument's socket.
5. Set up a test on the Micro-ohmmeter Model 6292, as instructed by
the instrument's user manual (located on the USB stick).
6. Clamp the probe on the ground conductor to be tested, then start the
test. The Model 6292 will display the measured ground conductor
current as the "Ig" reading on the instrument's LCD display panel:
In the preceding example, the Ig measurement provided by the
MR6292 probe is 8.0A.
8