10. Top Components
air gap
test port upper wash
overflow tray
10.1 Removing and reinstalling the USB port (Figure 38)
1. Remove top cover, left and right panels. (See Section 4. Removing and Replacing Panels.)
2. Remove chemical bracket. (See Section 7.6 Removing and replacing the chemical bracket.)
3. Remove 2 screws holding USB port to chassis.
4. Remove door fascia. (See Section 6.1 Removing and replacing the door fascia.)
5. Remove kickplate. (See Section 5.1 Removing and replacing the kickplate.)
6. Disconnect USB from the LCD controller board.
7. Cut cable ties on wiring harness and pull cable through to remove.
To reinstall:
1. Thread USB cable to LCD controller board and connect.
2. Fasten USB port to chassis.
3. Reinstall kickplate, door fascia, chemical bracket and panels.
Cable, USB – part # 01-113261S
upper wash
arm inlet
arm sensor
DC power
source
65
Figure 38
USB port