Removing And Reinstalling The Power Switch; Removing And Reinstalling The Usb Port - SciCan HYDRIM L110w G4 Service Manual

Instrument washer
Hide thumbs Also See for HYDRIM L110w G4:
Table of Contents

Advertisement

7. Right Side Components

7.8 Removing and reinstalling the power switch

1. Turn off the unit and disconnect the power.
2. Remove the screw above the LCD touchscreen.
3. Open the service panel.
4. Disconnect the wires from the power switch and press
down the locking tabs to remove, pushing the switch through the panel.
To reinstall, push the switch into position and reconnect wiring according to
Figure 30b. Be sure to pull out the locking tabs at the top left and bottom
right as shown in Figure 30a.
Rocker switch – part # 01-112024S

7.9 Removing and reinstalling the USB port

1. Turn unit off and disconnect the power.
2. Remove the screw above the touchscreen to tip open
the service panel.
3. Remove the retaining nuts from the threaded posts,
remove the bracket and disconnect the USB port from
the LCD touchscreen controller board.
To reinstall, reverse removal instructions.
Cable USB – part # 01-112398S
Pull tabs out
4
3
Figure 30a
USB port
60
5
24/25
Figure 30b
Figure 31
retaining nuts