SciCan HYDRIM C61w G4 Service Manual page 32

Instrument washer
Hide thumbs Also See for HYDRIM C61w G4:
Table of Contents

Advertisement

3 Diagnostics and Troubleshooting
Cycle Fault
CF 8 Cycle Fault
CF 9 Program
Timeout
CF 10 Drying System
Error
CF 13 Temperature
Validation Error
CF 14
CF 15 Chamber
Overflow
Effect
Problem
Drying aborted
Air temperature less
than a set point after
a timeout
Cycle interrupted
The unit is running
a cycle for more than
3h ±3 min.
Cycle interrupted
Air Dryer RPM not
zero when Dryer motor
should not be activated
Cycle interrupted
Water temperature rose
above the maximum
allowed temperature
Cycle interrupted
During the Prewash
phase the water
temperature in the
chamber is 5ºC higher
than the target for more
than 1sec
Cycle interrupted
Leak in the unit
32
Possible Causes
This is caused by a heater
malfunction:
Check air heater wire
harness for loose
contacts.
Check that the
heater element is not
interrupted.
Check I/O PCB air
heater relay output.
Defective PCB and/or
software failure
Replace Color LCD
controller.
Electronics – motor driver
failure (I/O board)
Check Dryer motor
wiring.
Verify that motor stops
when in non drying
phase.
Replace I/O board.
Replace dryer motor.
The temperature sensor out
of range:
Check temperature
sensor wires for loose
contacts.
Run cycle to monitor
that the water
temperature is below
96°C.
Replace sensor and
verify if the CF persist.
Replace I/O PCB.
The water reservoir overflow
switch was triggered:
Check the water
reservoir full switch.