8. Left Side Components
9. Disconnect the pump breather tube at the pump. (Figure 31b)
10. Disconnect the 2 hose clamps using pliers to slide them down
and out of the way. (Figure 31c)
11. To remove the air gap,
pull it up to clear
the water softener
connections.
To reinstall:
1. Apply soap (or silicone lubricant, if available) to the water softener
connections and push the air gap down into position. Ensure
the overflow outlet at the top of the air gap is properly aligned
with the tray.
2. Fasten the 2 screws at the left side of the air gap.
3. Reconnect tubing and tighten hose clamps.
4. Reconnect the pump breather tube to the pump and fasten
with a cable tie.
5. Reconnect the wire connections according to the color-coded
cable ties or your own markings.
6. Reinstall panels.
Backflow preventer – part # 01-113318S
breather tube
58
Figure 31b
hose clamps
Figure 31c
water softener
connections
Figure 31d