SciCan HYDRIM C61w G4 Service Manual page 25

Instrument washer
Hide thumbs Also See for HYDRIM C61w G4:
Table of Contents

Advertisement

2 Routine Procedures and Maintenance
• Inspect and clean sump filters (check sump for debris)
• Check wash arms for blockages and remove them for cleaning if required.
• Review error history
• Software upgrade (if necessary)
• Check individual component functionality. Go to the technician menu (enter access
code 7919) and select 'Diagnostic Tools' then select 'Component Tests'. From here
you can scroll through and check the functionality of the following components:
• Cooling fans
• Air gap pump
• RO valve (if fitted)
• Condenser valve (if fitted)
• Chamber heater
• Door latch
• Salt regeneration valve
• Dosing pump
• Dryer
• Hot water valve
• Cold water valve
• Air solenoid valve (if fitted)
• Waste pump
• Recirculation pump
• Check program selection
• Check dosing pump volume; dosing pump is pre-calibrated. Volume cannot
be adjusted.
• Check thermocouple calibration and adjust if required.
• Reset service cycle counter
• Clean machine
• Conduct electrical safety tests
Equipment and parts requirements for annual service
• Dryer filter (HEPA) (Part number 01-113277S)
• Main chamber seal (Part number 01-113300S)
• Lower door seal (Part number 01-113661S)
• Service Manual (Part number 95-113023)
• Electrical safety test equipment
• Water hardness test strips (Part number 01-108305S)
• Calibrated independent temperature measuring device
• 100ml graduated measuring cylinder
25

Hide quick links:

Advertisement

Table of Contents
loading