AGL7 Analog Genlock Module Troubleshooting

AGL7 Analog Genlock Module Troubleshooting

Equipment Required

Backing Up the Instrument Settings

3–14
This section contains instructions and procedures for troubleshooting the AGL7
Genlock module. These procedures will help you decide whether it is necessary to
replace your module.
If the TG700 mainframe does not boot up or if the display is not working, refer to
the TG700 Mainframe Troubleshooting section in this manual. (See page 1-24,
TG700 Mainframe Troubleshooting.)
The equipment listed in the following table is required for troubleshooting.

Table 3-3: Equipment required for AGL7 Analog Genlock module troubleshooting

Equipment description
Screwdriver handle
#1 Phillips tip
TG700 mainframe
Before you begin the troubleshooting procedures, create a preset to save your
instrument settings as follows:
1. Press the MODULE button until the mainframe main menu (TG700:
PRESET) appears.
2. Press the ENTER button to access the PRESET submenu.
3. Press the up (▲) or down (▼) arrow button to select SAVE.
4. Press the left (◄) or right (►) arrow button to select a preset number to save
instrument settings.
5. Press the ENTER button to save the current instrument settings to the new
preset.
After the problem is solved, recall the preset you created.
Minimum requirements/example
Accept Phillips-driver bits
Phillips-driver bit for #1 size screw heads
A TG700 mainframe that is functioning
correctly
TG700 TV Signal Generator Platform Service Manual