Operation; Making Measurements; Dc Measurement - AEMC MR561 User Manual

Ac/dc oscilloscope current probes
Table of Contents

Advertisement

4.1

Making Measurements

Connect the current probe to the proper input channel on the oscil-
loscope.
Begin with the least sensitive range on the current probe (1mV/A).
Select the 0.5V/Division on your oscilloscope.
Turn oscilloscope power on. Turn on the Model MR461/561. The
green LED should be on, the red LED should be off. If the green
indicator does not come on or goes off before the probe has oper-
ated for 10 minutes, it is necessary to replace the battery.
After approximately 10 minutes of operation, if none of the con-
trol buttons have been manipulated, power will automatically shut
off.
4.1.1

DC Measurement

Select DC position on the oscilloscope.
"Zero" the probe. With the probe disconnected from test samples
(no conductor in probe jaw window), press the auto zero button.
The red LED (%) comes on for approximately three seconds to
indicate that the probe is zeroing. If the red LED remains lit, this
indicates that zero has not been attained – repeat the zeroing
operation. Alternatively, you may "zero" with the oscilloscope.
For best accuracy, especially on low-level measurement, it is recom-
mended that you zero the probe before each measurement.
Clamp the probe around the conductor to be tested. The oscillo-
scope should now display the measured conductor current. A posi-
tive reading indicates current flowing in the direction of the arrow
located on the side of the jaw. A negative reading indicates current
flow in the opposite direction of the arrow. Multiply the conversion
ratio (1mV/A or 10mV/A) times the V/Division range on your oscil-
loscope to get the value of the current.
14
Shop for AEMC products online at:
CHAPTER 4

OPERATION

AC/DC Current Probe Models MR461 and MR561
ShopAEMC
www.
.ca
1.800.561.8187