Noise:
60A Range:
DC to 1kHz ≤ 8mV; DC to 5kHz: ≤ 12mV; 0.1Hz to 5kHz: ≤ 2mV
600A Range:
DC to 1kHz ≤ 1mV; DC to 5kHz: ≤ 1.5mV; 0.1Hz to 5kHz: ≤ 500µV
Rise and Fall Time:
60A Range: ≤ 100µs from 10 to 90% Vout
600A Range: ≤ 70µs from 10 to 90% Vout
3.2.2
MECHANICAL
Humidity Influence: 10 to 90% RH @ reference temperature ≤ 0.5%
Maximum Cable Diameter: One 1.18" (30mm) or two 0.95" (24mm) or
two bus bars 1.24 x 0.39" (31.5 x 10mm)
Dimensions (without zero knob): 8.8 x 3.82 x 1.73" (224 x 97 x 44mm)
Weight: 15oz (440g)
3.3
MR561 Specifications
3.3.1
ELECTRICAL
Measurement Range:
150A Range:
0.2 to 100A
AC
1500A Range:
0.5 to 1000A
AC
Output Signal:
10mV/A on 150A range; 1mV/A on 1500A range
Accuracy:
150A Range:
0.5 to 100A: 1.5% reading ± 0.5A;
100 to 150A
DC
1500A Range:
1 to 100A: 1.5% reading ± 1A; 100 to 800A: 2.5% reading ± 0.5A
800 to 1000A: 4% reading; 1000 to 1200A
10
Shop for AEMC products online at:
(150A Peak); 0.4 to 150A
(1400A Peak); 0.5 to 1500A
only: 2.5% reading
AC/DC Current Probe Models MR461 and MR561
ShopAEMC
www.
DC
DC
only: 4% reading
DC
.ca
1.800.561.8187