Phase Shift:
150A Range:
45 to 65Hz, 10 to 20A: ≤ 3.5°; 20 to 100A: ≤ 3°
1500A Range:
45 to 65Hz, 10 to 200A: ≤ 3°; 200 to 1000A: ≤ 2.5°
Overload: 1500A
Noise:
150A Range:
DC to 1kHz ≤ 8mV; DC to 5kHz: ≤ 12mV; 0.1Hz to 5kHz: ≤ 2mV
1500A Range:
DC to 1kHz ≤ 1mV; DC to 5kHz: ≤ 1.5mV; 0.1Hz to 5kHz: ≤ 500µV
Rise and Fall Time:
150A Range: ≤ 100µs from 10 to 90% Vout
1500A Range: ≤ 70µs from 10 to 90% Vout
3.3.2

MECHANICAL

Humidity Influence: 10 to 90% RH @ reference temperature ≤ 0.1%
Maximum Cable Diameter: One 1.6" (42mm) or two 0.98" (25.4mm) or
two bus bars 1.96 x 0.19" (50 x 5mm)
Dimensions (without zero knob):
9.31 x 3.82 x 1.73" (236.5 x 97 x 44mm)
Weight: 16oz (480g)
3.4
Common Specifications (MR461 & MR561)
3.4.1

ELECTRICAL

Frequency Range: DC to 10kHz @ -3 dB
Load Impedance: >100kΩ/100pF
Insertion Impedance: 0.39mΩ @ 50Hz, 58mΩ @1000Hz
Working Voltage: 600Vrms
Common Mode Voltage: 600Vrms
AC/DC Current Probe Models MR461 and MR561
Shop for AEMC products online at:
and 1000A
DC
AC
ShopAEMC
www.
continuous up to 1kHz
.ca
11
1.800.561.8187