Inserting a Cell

Your spectrophotometer is shipped with the standard single-cell holder which
you must install in the sample compartment. This cell holder accommodates
standard cells or flow cells. To insert a sample cell in the cell holder:
1 Move the locking lever to its up position.
3 Lock the sample cell in place by pushing the locking
lever back down.
Agilent Cary 8454 UV-Visible Spectroscopy System Operator's Manual
Good Measurement Practices
2 Insert the sample cell, making sure you orient it
correctly. The frosted (non-clear) sides of the
sample cell should not be in the path of the light
beam.
Utmost care should be taken when using small
volume flow cells and particularly any cells with less
than a 2 mm aperture. Here it is important that the
cells are properly centered in the light path. The cell
holder is designed for cells with 15 mm center
height. If applicable, these cells should not be
removed between the respective reference (blank)
measurement and the sample measurement.
4
Inserting a Cell
65