PERFORMANCE TEST
6.3.39 Z-MOD sensitivity
This test checks the sensitivity of the Z modulation facility.
test equipment:
Fluke 5500A mode: wavegen, wave square (Alternative: function generator PM5136)
T-piece
test set-up:
5500A
settings/procedure and requirements:
1 -
Apply a 1 kHz square-wave signal, duty cycle 50%, amplitude between 0 to +2.5V, to input
CH1. Adjust the generator to a signal of 2.5 Vpp with an offset of 1.25V.
2 -
Press the AUTOSET key.
3 -
Set MTB to 0.5 ms/div.
4 -
Set the trace of CH1 in mid position with the CH1 POS control.
5 -
Apply the same signal by means of the T-piece to the Z input (rear side).
6 -
Adjust TRACE INTENSITY so that only the bottom half of the square-wave is displayed: the
top half must be just invisible (0.5 ms light on; 0.5 ms light off).
7 -
Decrease the input signal to 0.5 Vpp with an offset of 0.25V.
8 -
Set CH1 to 0.5 V/div.
9 -
Verify that the complete square-wave is visible.
OSCILLOSCOPE UNDER TEST
6 - 3 3