Fluke PM3082 Service Manual page 29

Analog oscilloscopes
Table of Contents

Advertisement

2 - 1 8
CHARACTERISTICS
The PM3082 and PM3084, including standard accessories, conform with the EEC Directive 89/336
for EMI immunity, as defined by IEC 801-3, with the addition of the following tables.
Frequency range:
10kHz.. 25 MHz
Stand alone
With PM9010/091
Frequency range:
25 MHz ... 1 GHz
Stand alone
With PM9010/091
Frequency range:
10kHz.. 25 MHz
Stand alone
With PM9010/091
Frequency range:
25 MHz ... 1 GHz
Stand alone
With PM9010/091
For conditions not specified in tables 1 - 4, a susceptibility effect of more than 10 % is possible.
SPECIFICATION
Table 1.
Susceptibility: no visible disturbance
E = < 0.1 V/m
2 mV/div... 5 V/div
2 mV/div... 5 V/div
Table 2.
Susceptibility: no visible disturbance
E = < 0.1 V/m
2 mV/div... 5 V/div
2 mV/div... 5 V/div
Table 3.
Susceptibility: disturbance less than 10 % of full scale
E = < 0.1 V/m
N/A
N/A
Table 4.
Susceptibility: disturbance less than 10 % of full scale
E = < 0.1V/m
N/A
N/A
ADDITIONAL INFORMATION
E = 1 V/m
2 mV/div ... 5 V/div
10 mV/div ... 5 V/div
E = 1 V/m
5 mV/div ... 5 V/div
50 mV/div ... 5 V/div
E = 1 V/m
N/A
2 mV/div ... 5 mV/div
E = 1 V/m
2 mV/div
10 mV/div... 20
mV/div
CHARACTERISTICS
E = 3 V/m
2 mV/div ... 5 V/div
20 mV/div ... 5 V/div
E = 3 V/m
10 mV/div ...5 V/div
200 mV/div ... 5 V/div
E = 3 V/m
N/A
5 mV/div ... 10 mV/div
E = 3 V/m
2 mV/div ... 5 mV/div
50 mV/div.... 100
mV/div