Product Specifications —
This chapter gives the product specifications for the SurePower Single Bay Charger.
Size
Weight
Equipment Type
Power Requirements
Power Consumption
Output
Maximum Thermal
Output
Design Standards
Temperature
Humidity
Safety Specification
Altitude of Operation
Ingress Protection
9650-000272-05 Rev. B
SurePower Single
8.8 (Length) x 6.8 (Width) x 5.0 (Height) in
22.4 (Length) x 17.3 (Width) x 12.7 (Height) cm
4 lb (1.8 kg) without batteries
IEC Class 1
100-240 Vac, 50/60 Hz, 125VA
120 VA
Voltage: +6.0 to 13.5 V dc.
Charging Current: 0 to +3.0 A dc.
Discharging Current: 0 to 2.5 A dc.
70 BTU/HR
Designed to meet or exceed IEC 60601-1 standards for
medical equipment safety
Operation:
32°F to 122°F
0°C to 50°C
Storage and Transportation:
-40°F to 158°F
-40°C to 70°C
5% to 95% (relative humidity - non condensing).
Relative humidity for operation at temperatures
40°C - 50°C is limited to a maximum of 57%.
Stationary, Class, Type
-300 ft through 13,123 ft (below and above sea level)
-91 m through 4,000 m (below and above sea level)
IPX2 (equipment is protected against harmful effects from
water ingress of vertically falling water drops when
enclosure is tilted up to 15°)