Surepower Charger Station; Product Specifications - ZOLL SurePower Operator's Manual

Charger station
Hide thumbs Also See for SurePower:
Table of Contents

Advertisement

This chapter gives the product specifications for the ZOLL
Size
Weight
Equipment Type
Power Requirements
Power Consumption
Output per Bay
Maximum Thermal
Output
Design Standards
Temperature
Humidity
Safety Specification
9650-0535-01 Rev. C

Product Specifications --

SurePower Charger Station

16 (Length) x 11 (Width) x 6 (Height) in
40.6 (Length) x 27.9 (Width) x 15.2 (Height) cm
15 lb (6.8 kg) without batteries
IEC Class I
100-240 Vrms, 50/60 Hz
120 Vrms, 400 Hz
200 VA maximum.
Voltage: +6 to 16.8 V dc
Charging Current: 0 to +3.00 A dc
Discharging Current: 0 to -3.00 A dc
682 BTU
Designed to meet or exceed UL 60601-1, EN 60601-1,
and CSA 22.2 standards for medical equipment safety
Operation:
32°F to 122°F
0°C to 50°C
Optimal battery charging:
59°F to 95°F
15°C to 35°C
Storage and transportation:
-40°F to 158°F
-40°C to 70°C
5% to 95% (relative humidity - non condensing).
Stationary, Class I, Type B
SurePower Charger Station Operator's Guide