This chapter gives the product specifications for the ZOLL
Size
Weight
Equipment Type
Power Requirements
Power Consumption
Output per Bay
Maximum Thermal
Output
Design Standards
Temperature
Humidity
Safety Specification
9650-0535-01 Rev. C
Product Specifications --
SurePower Charger Station
16 (Length) x 11 (Width) x 6 (Height) in
40.6 (Length) x 27.9 (Width) x 15.2 (Height) cm
15 lb (6.8 kg) without batteries
IEC Class I
100-240 Vrms, 50/60 Hz
120 Vrms, 400 Hz
200 VA maximum.
Voltage: +6 to 16.8 V dc
Charging Current: 0 to +3.00 A dc
Discharging Current: 0 to -3.00 A dc
682 BTU
Designed to meet or exceed UL 60601-1, EN 60601-1,
and CSA 22.2 standards for medical equipment safety
Operation:
32°F to 122°F
0°C to 50°C
Optimal battery charging:
59°F to 95°F
15°C to 35°C
Storage and transportation:
-40°F to 158°F
-40°C to 70°C
5% to 95% (relative humidity - non condensing).
Stationary, Class I, Type B
SurePower Charger Station Operator's Guide
Chapter 6
SurePower Charger Station
:
6-1