ZOLL Base PowerCharger 4x4 Operator's Manual
ZOLL Base PowerCharger 4x4 Operator's Manual

ZOLL Base PowerCharger 4x4 Operator's Manual

Hide thumbs Also See for Base PowerCharger 4x4:

Advertisement

Quick Links

ZOLL Base PowerCharger
4x4
Operator's Manual
Part Number 9650-0054 Rev. G

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Base PowerCharger 4x4 and is the answer not in the manual?

Questions and answers

Summary of Contents for ZOLL Base PowerCharger 4x4

  • Page 1 ZOLL Base PowerCharger Operator’s Manual Part Number 9650-0054 Rev. G...
  • Page 2 Outside of North America, consult the ZOLL authorized representative. In order to maintain this warranty, the instructions and procedures contained in this guide must be strictly followed. For additional information, please call the ZOLL Technical Service Department 1-800-348-9011 in North America. (International customers: 978-421-9655).
  • Page 3: Table Of Contents

    Table of Contents 6HFWLRQ  *HQHUDO ,QIRUPDWLRQ   3URGXFW 'HVFULSWLRQ  +RZ WR 8VH 7KLV 0DQXDO  6HFWLRQ 'HVFULSWLRQV   0DQXDO 8SGDWHV  8QSDFNLQJ  )'$ 5HJXODWLRQV   :DUUDQW\ 86 2QO\   8VH RI 1RQ6WDQGDUG $FFHVVRULHV ZLWK =2// (TXLSPHQW   :DUQLQJV DQG &DXWLRQV   6SHFLILFDWLRQV  6\PEROV 8VHG RQ WKH (TXLSPHQW  ...
  • Page 4 ZOLL Base PowerCharger Operator’s Manual (This page intentionally left blank.)
  • Page 5 3URGXFW 'HVFULSWLRQ The ZOLL Base PowerCharger (with QuickCharge or AutoTest) is a battery charger and testing system designed for management of PD 4410/Smart/Smart Ready Battery and all XL Battery packs used in ZOLL Medical Corporation resuscitation devices. The ZOLL Base PowerCharger (with QuickCharge or AutoTest) provides four battery charging/testing compartments.
  • Page 6: 6Hfwlrq  *Hqhudo ,Qirupdwlrq

    If you purchased your product more than three years QuickCharge relies on the user to manually initiate the ago, contact ZOLL to determine whether updates to your battery test cycle at recommended intervals (see documentation are available. For information relating to “Manual Battery Testing Procedure”...
  • Page 7: Duudqw\ 86 2Qo

    Corporation for any breach of any warranty related to the equipment supplied hereunder. (e) Limitation of Liability: ZOLL shall not in any event be liable to Purchaser, nor shall Purchaser recover, for special, incidental or consequential damages resulting from any breach of...
  • Page 8: Duqlqjv Dqg &Dxwlrqv

    • indicator is lit and the FAULT light is off. Failure to do so may result in the use of a fully depleted battery. Only charge XL Battery packs in a ZOLL Base • PowerCharger marked “XL Battery Ready”. Failure to do so will reduce battery capacity and...
  • Page 9 5% to 95% (relative humidity - non FAULT Illuminates when battery capacity test or condensing). battery compartment test fails. Separate Capacity Test At a minimum, test batteries every indicator light for each battery three (3) months. Refer to ZOLL’s Battery compartment. Management Program brochure.
  • Page 10 Chelmsford, MA 01824-4105 Fusible Link Attn: Technical Service Department Loaner instruments are available for use while repairs Protective (earth) ground terminal are being completed by ZOLL's Technical Service Department. International Customers DANGER High Voltage present Should the ZOLL Base PowerCharger...
  • Page 11 QuickCharge Refer to the Operating Controls and Indicators section on the next page for a description of each number. 1RWH In this section, all references to ZOLL Base PowerCharger refer to QuickCharge unless otherwise specified. If your device does not indicate QuickCharge or AutoTest on the label (in the upper right corner), refer to the...
  • Page 12: 2Shudwlqj &Rqwurov Dqg ,Qglfdwruv

    The ZOLL PD 4410/Smart/Smart Ready Battery may be directly above. charged in any of the four battery compartments. XL Batteries should only be charged in a ZOLL Base  )$8/7 PowerCharger labeled “XL Battery Ready”. Each...
  • Page 13: 0Dqxdo %Dwwhu\ 7Hvwlqj 3Urfhgxuh

    To test a battery: CHARGER ON BATT. READY FAULT 1. Check to see that the ZOLL Base PowerCharger is connected to live AC mains. The POWER indicator TEST TESTING light should be lit.
  • Page 14: Dwwhu\ &Rpsduwphqw 7Hvwlqj 3Urfhgxuh

    In TEST TESTING addition, the ZOLL Base PowerCharger can test each battery compartment for defective operation through the 4. If the battery passes the test, the BATT. READY light simple press of a button.
  • Page 15: Dxow &Rghv Dqg 'Hilqlwlrqv

    Operation (QuickCharge) )DXOW &RGHV DQG 'HILQLWLRQV When the FAULT indicator light illuminates, press and hold the TEST button. The illumination of the following indicator lights together represent a specific defect, identified under the Fault Code column. &+$5*(5 21 %$77 5($'< )$8/7 7(67,1* )DXOW &RGH...
  • Page 16: 4Xlfn 5Hihuhqfh &Kduw Iru =2// %Dvh 3Rzhu&Kdujhu

    Manual Test is complete and the battery is NOT usable. 1RWH If a battery is removed when the FAULT light is illuminated, the ZOLL Base PowerCharger will beep to indicate the battery has been removed and should NOT be used.
  • Page 17 AutoTest Refer to the Operating Controls and Indicators section on the next page for a description of each number. 1RWH In this section, all references to ZOLL Base PowerCharger refer to AutoTest unless otherwise specified. If your device does not indicate QuickCharge or AutoTest on the label (in the upper right corner), refer to the...
  • Page 18: 2Shudwlqj &Rqwurov Dqg ,Qglfdwruv

    Also, illuminates when a battery charged in any of the four battery compartments. compartment or any other failure occurs. XL Batteries should only be charged in a ZOLL Base PowerCharger labeled “XL Battery Ready”. Each battery charge compartment operates independently of...
  • Page 19: 0Dqxdo %Dwwhu\ 7Hvwlqj 3Urfhgxuh

    ) for approximately not affect any other charge or testing cycles in progress. 1.5 hours in Monitor mode. To test a battery: 1. Check to see that the ZOLL Base PowerCharger CHARGER ON BATT. READY FAULT is connected to live AC mains. The POWER light should be lit.
  • Page 20: Dwwhu\ &Rpsduwphqw 7Hvwlqj 3Urfhgxuh

    The ZOLL Base PowerCharger has four battery compartments that are able to test and charge batteries simultaneously and independently of each other. In addition, the ZOLL Base PowerCharger can test each battery compartment for defective operation through the simple press of a button.
  • Page 21: Dxow &Rghv Dqg 'Hilqlwlrqv

    Operation (AutoTest) )DXOW &RGHV DQG 'HILQLWLRQV When the FAULT indicator light illuminates, press and hold the TEST button. The illumination of the following indicator lights together represent a specific defect, identified under the Fault Code column. &+$5*(5 21 %$77 5($'< )$8/7 7(67,1* )DXOW &RGH...
  • Page 22: 4Xlfn 5Hihuhqfh &Kduw Iru =2// %Dvh 3Rzhu&Kdujhu

    Manual Test is complete and the battery is NOT usable. 1RWH If a battery is removed when the FAULT light is illuminated, the ZOLL Base PowerCharger will beep to indicate the battery has been removed and should NOT be used.
  • Page 23: Qvshfwlrq

    Clean the ZOLL Base PowerCharger with a soft cloth, mild soap and water. Do not immerse any part of the ZOLL Base PowerCharger in water. Do not use alcohol or ketones (MEK, acetone, etc.). Do not autoclave the ZOLL Base...
  • Page 24: 7Urxeohvkrrwlqj

    If trouble persists after consulting this guide, contact the ZOLL Technical Service Department. 1. POWER indicator light does not illuminate. Check that unit is plugged into an appropriate •...
  • Page 25: 2// %Dwwhu\ 3Dfnv

    Section 5 Battery Management =2// %DWWHU\ 3DFNV Safe, reliable use of all ZOLL devices requires a well-designed battery management program to ensure ZOLL PD 4410/Smart/Smart Ready Battery and XL that adequate battery power is always available. Battery Packs are five-cell assemblies of sealed lead-...
  • Page 26 ZOLL Base PowerCharger Operator’s Manual (This page intentionally left blank.)
  • Page 28 + 61-2-432-922-26 Toll Free: (866) 442-1011 Telefax: + 61-2-432-922-26 Telefax: (905) 629-0575 Latin America Regional Office ZOLL Medical Australia Pty. Ltd Unit 4, 12 Chaplin Dr. 6468 NW 75 Lane Cove Parkland, FL 33067 NSW 2066 Australia Telephone: + 954-345-4224...

Table of Contents