Instructions for use PROPHYflex 3 - 2018 for Sirona connection - 1.006.9920, - 1.006.9927
2 Safety | 2.7 Service and repair
▶ Clean the suction hoses of the treatment centre after every application.
- For this purpose, aspirate approx. 200 ml of water with the tube to be cleaned.
- Make sure that the sliders at the cannula holders of the suction tubes are closed.
▶ Alternatively, use PROPHYflex Powder.

2.7 Service and repair

Repairs, servicing and safety checks may only be performed by trained service per-
sonnel. The following persons are authorised to do this:
▪ Service technicians of KaVo branches after the appropriate product training
▪ Service technicians of KaVo authorised dealers after the appropriate product train-
ing
Observe all the following items during servicing work:
▶ Have the service and testing tasks carried out in accordance with the Medical
Product Operator Ordinance.
▶ After servicing, interventions, and repairs of the device and before re-use, the
device must be subjected to safety checks by the service personnel.
▶ Following expiry of the warranty, check the cutter/grinder holding system and
quick stop once a year.
▶ KaVo recommends specifying in-house service intervals where the medical device
is brought to a professional shop for cleaning, servicing and a function check.
Define the service interval depending on the frequency of use.
Cleansers and disinfectants can damage the plastic housing leading to hairline cracks
and other damage which can ultimately lead to hazards.
▶ Have a safety check done on the product every 2 years. For this purpose, send
the product to the KaVo Customer Service Centre Warthausen or a test centre ap-
proved by KaVo.

2.8 Protective equipment

PROPHYflex Powder and other powders may get in the eyes or mouth of the user or
patient during treatment.
▶ Both patient and user have to wear protective goggles during the treatment.
▶ KaVo recommends the use of dust extraction equipment and mouth protection
during the treatment with PROPHYflex.

2.9 Property damage

The use of PROPHYflex Powder and other powders can cause scratches on parts
and/or products with sensitive surfaces. Wiping sensitive surfaces with a cloth can
cause scratches.
▶ Remove fine powder deposits with a dust extraction device.
▶ Use running tap water to rinse residual powder from liquid-insensitive parts.