Error Messages - Hoffrichter CPAP User Manual

Respiratory therapy device
Hide thumbs Also See for CPAP:
Table of Contents

Advertisement

.!00 E

ERROR MESSAGES


..+.
Two different types of error can be distinguished: errors with data loss and
errors without data loss.
The display "Exx" denotes errors in which data may have been lost. The display
"Wxx" denotes errors in which data has not been lost.
To hide an error message, press the enter key
If error messages occur frequently, we recommend that the device is
examined by a service technician.
The following table shows all error codes that can occur in connection with
the use of the SD card.
Error
Meaning
code
E00
Writing data has failed. Possible
causes are power failure, removal of
card during writing or a faulty SD card.
E01
The SD card is faulty.
E02
File system or folder structure is faulty
or cannot be read.
E03
New fi les cannot be created or the
SD card is full.
E04
The device cannot adopt the ther-
apy data.
E05
The SD card was removed.
The SD card has fallen out of the card
slot.
E06
The SD card or the device is faulty.
W00
The device is attempting to repair the
SD card. But you have inserted a new
SD card into the device.
W05/06 see error code E05/E06
62
Using an SD card
H
;=
.
Remedy
The SD card is probably faulty. Remove
the SD card and then insert it into
the device again. If the error contin-
ues to occur, send the card back to
the physician.
Obtain a new SD card from your phy-
sician or specialist dealer.
The SD card must be formatted by
the physician using TRENDset.
The SD card must be formatted by
the physician using TRENDset.
The SD card must be returned to the
physician to check the therapy.
Insert the SD card again.
If this occurs frequently, the device
should be examined by a service
technician.
Change SD card. If this occurs fre-
quently, the device should be exam-
ined by a service technician.
Confi rm with key. The old SD card is
defective. Continue to use the new
SD card.
see error code E05/E06

Advertisement