CALIBRATION

Calibration guarantees an optimum performance of the automatic func-
tion and mask test. It must be performed when:
the device is used initially,
the mask has been changed,
there is a change within the therapy system. If a bacterial fi lter or a
humidifi er is added/removed the calibration has to be redone.
Ex factory, the point 2 is pre-calibrated for using with a Cirri Comfort nasal
mask (size M) without humidifi er and without bacterial fi lter.
Calibration is only possible during mains operation not in battery mode.
Calibrating the device
1. Connect the mask, the therapy tube and, if applicable, the humidifi er or
bacterial fi lter to the device.
Ensure that the air and exhalation openings in the mask are clear and
the tube is not kinked. Do not put the mask on.
2. Press the enter key
3. Use the selection keys
4. Press the enter key
5.
appears on the display. Press the enter key
Start
press the key within 5 s, the display jumps back (see point 3).
6. The blower starts and calibration begins.
on the display. Calibration takes 20 - 70 s. The blower switches off after
calibration is fi nished.
7. If calibration has been successful,
and if it has failed,
8. In both cases press the enter key
The old values will be retained if calibration
has failed. Start calibration again.
50
Device functions
for 1 second.
and
to select
.
Successful
appears on the display.
Failed
Calibration
. If you do not
Calibration active
appears on the display
.
.
appears